Filmless evaluation of the mechanical accuracy of the isocenter in stereotactic radiotherapy.

Peter Geyer1, Hilbert Blank, Carsten Evers, Thomas Leichtner, Horst Alheit.   

Abstract

BACKGROUND AND
PURPOSE: The Winston-Lutz test verifies the mechanical accuracy of the isocenter in stereotactic radiotherapy. A lead ball inside a small beam is exposed to film applying different combinations of the gantry angle and the table angle. The increasing replacement of films by digital images requires alternative imaging methods. The suitability of two different electronic portal imaging systems and of a system based on digital luminescence radiography was investigated.
MATERIAL AND METHODS: The imaging systems included the portal imaging devices BEAMVIEW PLUS and OPTIVUE1000 (both Siemens Medical Solutions, Erlangen, Germany) and the luminescence system KODAK ACR 2000 RT (Eastman Kodak Comp., Rochester, NY, USA). 6-MV photons from the linear accelerators PRIMUS and ONCOR (both Siemens Medical Solutions) were applied. First, only the small beam covering the lead ball was exposed. Second, an additional bigger open beam part in a certain distance to the small beam was applied.
RESULTS: For all three investigated imaging systems, which are using preprocessing imaging software, only for the beam arrangement with additional open beam parts, the lead ball could be detected inside the small beam. Only for the application of a dosimetric software tool to the luminescence system, the metal ball inside the small beam became visible without an additional open beam part.
CONCLUSION: Applying the proposed beam arrangements, the Winston-Lutz test can be done by digital and filmless imaging systems, thereby saving time as well.

Mesh:

Year:  2007        PMID: 17294111     DOI: 10.1007/s00066-007-1659-4

Source DB:  PubMed          Journal:  Strahlenther Onkol        ISSN: 0179-7158            Impact factor:   3.621


  7 in total

1.  On the accuracy of isocenter verification with kV imaging in stereotactic radiosurgery.

Authors:  Rolf Wiehle; Hans-Jürgen Koth; Norbert Nanko; Anca-Ligia Grosu; Norbert Hodapp
Journal:  Strahlenther Onkol       Date:  2009-05-15       Impact factor: 3.621

2.  Abstracts of the 1st International Conference on the Clinical Use of Tomotherapy. October 17-18, 2008. Munich, Germany.

Authors: 
Journal:  Strahlenther Onkol       Date:  2009-01       Impact factor: 3.621

3.  Comparison of stereotactic radiosurgery and fractionated stereotactic radiotherapy of acoustic neurinomas according to 3-D tumor volume shrinkage and quality of life.

Authors:  Martin Henzel; Klaus Hamm; Helmut Sitter; Markus W Gross; Gunnar Surber; Gabriele Kleinert; Rita Engenhart-Cabillic
Journal:  Strahlenther Onkol       Date:  2009-09-12       Impact factor: 3.621

4.  Hypofractionated stereotactic reirradiation of recurrent glioblastomas : a beneficial treatment option after high-dose radiotherapy?

Authors:  Emmanouil Fokas; Ulrich Wacker; Markus W Gross; Martin Henzel; Elitsa Encheva; Rita Engenhart-Cabillic
Journal:  Strahlenther Onkol       Date:  2009-04-16       Impact factor: 3.621

5.  On the selection of gantry and collimator angles for isocenter localization using Winston-Lutz tests.

Authors:  Weiliang Du; Jennifer L Johnson; Wei Jiang; Rajat J Kudchadker
Journal:  J Appl Clin Med Phys       Date:  2016-01-08       Impact factor: 2.102

Review 6.  Isocenter verification for linac-based stereotactic radiation therapy: review of principles and techniques.

Authors:  Pejman Rowshanfarzad; Mahsheed Sabet; Daryl J O'Connor; Peter B Greer
Journal:  J Appl Clin Med Phys       Date:  2011-11-15       Impact factor: 2.102

7.  Independent evaluation of the effectiveness of IsoCal in improving image center accuracy on Varian TrueBeam and Clinac machines.

Authors:  Weiliang Du; Song Gao; Wei Jiang; Rajat J Kudchadker
Journal:  J Appl Clin Med Phys       Date:  2018-06-29       Impact factor: 2.102

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.