| Literature DB >> 17292986 |
Felicita Andreotti, Alberto Lavorgna, Giulio Coluzzi, Claudio Vincenzoni, Antonio G Rebuzzi, Italo Porto, Filippo Crea.
Abstract
We report an unusual case of breakage and embolisation of a femoral access closure device (Angio-Seal), which caused acute leg ischemia after percutaneous coronary intervention in a diabetic patient with documented peripheral vascular disease. Embolectomy from the superficial femoral artery revealed the polymer anchor split in two parts and embedded within a large (25x5 mm) thrombus.Entities:
Mesh:
Year: 2007 PMID: 17292986 DOI: 10.1016/j.ijcard.2006.07.027
Source DB: PubMed Journal: Int J Cardiol ISSN: 0167-5273 Impact factor: 4.164