Determination of catecholamines by flow-injection analysis and high-performance liquid chromatography with chemiluminescence detection.

Edyta Nalewajko1, Aneta Wiszowata, Anatol Kojło.   

Abstract

A chemiluminescence (CL) detection of catecholamines [norepinephrine (NE), epinephrine (E), dopamine (DA) and L-dopa (LD)] is described for the flow-injection (FI) and high-performance liquid chromatographic (HPLC) determination of these compounds. The detection method is based on the inhibition effect of catecholamines (CAs) on the CL reaction of luminol with iodine in the alkaline medium. The proposed FI method allows the determination of CAs in pharmaceutical preparations for the purpose of drug quality control. The calibration curves show good linearity in the concentration range of: 1.1-20.0 microg l(-1) (NE), 0.5-5.0 microg l(-1) (E), 0.6-9.0 microg l(-1) (DA) and 0.6-10.0 microg l(-1) (LD). The limits of detection (defined as a signal-to-noise ratio of 3) are: 0.34 microg l(-1) (NE), 0.15 microg l(-1) (E) and 0.18 microg l(-1) (DA, LD). The HPLC procedure was successfully applied for the determination of catecholamines (NE, E, DA) in human urine after solid-phase extraction (SPE). In a simple run time CAs can be determined in 20 min. The chromatographic linear ranges are: 5.0-72.0 microg l(-1) (NE), 5.0-48.0 microg l(-1) (E) and 5.0-96.0 microg l(-1) (DA). The limits of detection for three urinary CAs are: 0.71 microg l(-1) (NE), 0.26 microg l(-1) (E) and 0.73 microg l(-1) (DA).

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 17289328     DOI: 10.1016/j.jpba.2006.12.021

Source DB:  PubMed          Journal:  J Pharm Biomed Anal        ISSN: 0731-7085            Impact factor:   3.935


  5 in total

Review 1.  Luminol-based chemiluminescent signals: clinical and non-clinical application and future uses.

Authors:  Parvez Khan; Danish Idrees; Michael A Moxley; John A Corbett; Faizan Ahmad; Guido von Figura; William S Sly; Abdul Waheed; Md Imtaiyaz Hassan
Journal:  Appl Biochem Biotechnol       Date:  2014-04-22       Impact factor: 2.926

Review 2.  Quantitative detection of neurotransmitter using aptamer: From diagnosis to therapeutics.

Authors:  Koel Sinha; Chitrangada DAS Mukhopadhyay
Journal:  J Biosci       Date:  2020       Impact factor: 1.826

3.  Determination of free catecholamines in urine by tandem affinity/ion-pair chromatography and flow injection analysis.

Authors:  David H Thomas; Jay D Taylor; Omar S Barnaby; David S Hage
Journal:  Clin Chim Acta       Date:  2008-08-22       Impact factor: 3.786

4.  Spectrophotometric Determination of Dopamine in Bulk and Dosage Forms Using 2,4-Dinitrophenylhydrazine.

Authors:  Mai Abed Alrahman Ramadan; Ihab Almasri; Ghada Khayal
Journal:  Turk J Pharm Sci       Date:  2020-12-23

Review 5.  Analysis of Catecholamines and Pterins in Inborn Errors of Monoamine Neurotransmitter Metabolism-From Past to Future.

Authors:  Sabine Jung-Klawitter; Oya Kuseyri Hübschmann
Journal:  Cells       Date:  2019-08-09       Impact factor: 6.600

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.