Literature DB >> 17285420

Bilateral tension pneumothoraces following jet ventilation via an airway exchange catheter.

Chris Nunn1, Joshua Uffman, Sanjay M Bhananker.   

Abstract

We report a case involving a 55-year-old man who had a recent resection of tracheal carcinoma and tracheal reanastomosis. He subsequently developed tracheomalacia and anastomotic dehiscence requiring airway stenting via an armored endotracheal tube (ETT). Placement of the armored ETT was technically difficult. It required insertion of an airway exchange catheter through the tracheal stoma to oxygenate, ventilate, and serve as a guide for ETT placement through the tracheotomy and across the dehiscence. During transtracheal jet ventilation our patient developed bilateral tension pneumothoraces requiring cardiopulmonary resuscitation and chest tube placement. The patient was quickly recovered, stabilized, and later discharged after a prolonged intensive care unit (ICU) course. We review the recommendations for jet ventilation via airway exchange catheters, common problems during this technique, and potential methods for avoiding these problems. The risk of barotrauma and pneumothoraces during jet ventilation via an airway exchange catheter should be kept in mind.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 17285420     DOI: 10.1007/s00540-006-0463-0

Source DB:  PubMed          Journal:  J Anesth        ISSN: 0913-8668            Impact factor:   2.078


  15 in total

1.  Airway exchange catheters: simple concept, potentially great danger.

Authors:  J L Benumof
Journal:  Anesthesiology       Date:  1999-08       Impact factor: 7.892

2.  Endobronchial rupture from endotracheal reintubation with an endotracheal tube guide.

Authors:  P A Seitz; N Gravenstein
Journal:  J Clin Anesth       Date:  1989       Impact factor: 9.452

3.  Quantification of the jet function of a jet stylet.

Authors:  S D Gaughan; J L Benumof; G T Ozaki
Journal:  Anesth Analg       Date:  1992-04       Impact factor: 5.108

4.  [Intraoperative exchange of the endotracheal tube using the Cook C-CAE airway exchange catheter].

Authors:  P Lambotte; H Menu; T Guermouche; E Boufflers; J Ferri; R Krivosic-Horber
Journal:  Ann Fr Anesth Reanim       Date:  1998

5.  The effective tracheal diameter that causes air trapping during jet ventilation.

Authors:  R Dworkin; J L Benumof; R Benumof; T G Karagianes
Journal:  J Cardiothorac Anesth       Date:  1990-12

6.  The use of an endotracheal ventilation catheter in the management of difficult extubations.

Authors:  R M Cooper
Journal:  Can J Anaesth       Date:  1996-01       Impact factor: 5.063

Review 7.  Airway exchange catheters for safe extubation: the clinical and scientific details that make the concept work.

Authors:  J L Benumof
Journal:  Chest       Date:  1997-06       Impact factor: 9.410

8.  A pressure curve monitor for intratracheal jet ventilation.

Authors:  G A Baer; J A Malmivuo; P P Talonen
Journal:  J Biomed Eng       Date:  1983-04

Review 9.  Studies in transtracheal ventilation catheters.

Authors:  C A Rone; E G Pavlin; C W Cummings; E A Weymuller
Journal:  Laryngoscope       Date:  1982-11       Impact factor: 3.325

10.  Adverse respiratory events infrequently leading to malpractice suits. A closed claims analysis.

Authors:  F W Cheney; K L Posner; R A Caplan
Journal:  Anesthesiology       Date:  1991-12       Impact factor: 7.892

View more
  3 in total

1.  Pneumothorax, an underappreciated complication with an airway exchange catheter.

Authors:  Ali M Rashid; Charles Williams; Jason Noble; Omar M Rashid; Kazuaki Takabe; Rahul J Anand
Journal:  J Thorac Dis       Date:  2012-12       Impact factor: 2.895

2.  Hypoxic Respiratory Failure Further Complicated During Airway Exchange Catheter Placement.

Authors:  Osman Ali; Avelino C Verceles; Matthew L Barrett; Christopher M Franklin; Imran Khan; Negar Naderi; Michael T McCurdy
Journal:  J Crit Care Med (Targu Mures)       Date:  2020-08-11

3.  Airway tube exchanger techniques in morbidly obese patients.

Authors:  Danai Udomtecha
Journal:  Anesthesiol Res Pract       Date:  2012-02-09
  3 in total