Laser fluorescence measurements compared to electrical resistance of residual dentine in excavated cavities in vivo.

F Krause1, A Braun, J Eberhard, S Jepsen.   

Abstract

It has been suggested that laser fluorescence close to the dental pulp shows higher values than more distant measurements. The aim of this study was to assess fluorescence on the cavity floor and to correlate these measurements with electrical resistance as a measure of residual dentine thickness. Thirty carious lesions were excavated with a bur. The endpoint of caries removal was determined by visual, tactile and auditory means. Fluorescence was measured with the Diagnodent device and with the fluorescence feedback system of a therapeutic Er:YAG laser. Electrical resistance of the residual dentine (Prepometer units, PU) was measured with a proprietary device. Significant differences were observed between the fluorescence systems (p < 0.05). For Diagnodent, a decrease of electrical resistance of the residual dentine resulted in an increase of the fluorescence values of 2.99 units/PU (95% CI = 2.00-3.97) and an increase of 0.30 units/PU (95% CI = 0.19-0.40) for the fluorescence feedback system. For zero electrical resistance, a predicted maximum value of 51.5 units (95% CI = 41.3-61.7) was calculated for the Diagnodent and 5.1 units (95% CI = 4.1-6.2) for the feedback system. The study indicates that using the suggested detection cut-off with the Diagnodent device may be not suited to assess the endpoint of caries removal close to the dental pulp. Employing the Key Laser III, values up to 6 units might be caused solely by close proximity to the pulp, which should be considered when caries removal by laser is controlled by laser fluorescence feedback. Copyright 2007 S. Karger AG, Basel.

Entities:  

Mesh:

Year:  2007        PMID: 17284915     DOI: 10.1159/000098047

Source DB:  PubMed          Journal:  Caries Res        ISSN: 0008-6568            Impact factor:   4.056


  16 in total

1.  The influence of zero value subtraction on the performance of a new laser fluorescence device for approximal caries detection.

Authors:  Jonas Almeida Rodrigues; Jonas de Almeida Rodrigues; Isabel Hug; Adrian Lussi
Journal:  Lasers Med Sci       Date:  2008-02-29       Impact factor: 3.161

2.  In vivo assessment of caries excavation with a fluorescence camera compared to direct bacteriological sampling and quantitative analysis using flow cytometry.

Authors:  Richard Stoll; Beata Urban-Klein; Paul Giacomin; Alex Loukas; Anahita Jablonski-Momeni
Journal:  Lasers Med Sci       Date:  2013-10-19       Impact factor: 3.161

3.  Fluorescence of primary dentine after chemomechanical and conventional rotary excavation.

Authors:  F N P Corrêa; R O Rocha; F Z M Soares; L E Rodrigues-Filho; C R M Rodrigues
Journal:  Eur Arch Paediatr Dent       Date:  2008-09

4.  Caries detection under composite restorations by laser fluorescence and digital radiography.

Authors:  Suwadee Kositbowornchai; Chotikpanich Sukanya; Tangkittikasem Tidarat; Thongsombat Chanoggarn
Journal:  Clin Oral Investig       Date:  2012-12-16       Impact factor: 3.573

Review 5.  Evaluation of Accuracy of DIAGNOdent in Diagnosis of Primary and Secondary Caries in Comparison to Conventional Methods.

Authors:  Hanieh Nokhbatolfoghahaie; Marzieh Alikhasi; Nasim Chiniforush; Farzaneh Khoei; Nassimeh Safavi; Behnoush Yaghoub Zadeh
Journal:  J Lasers Med Sci       Date:  2013

6.  Cavity size difference after caries removal by a fluorescence-controlled Er:YAG laser and by conventional bur treatment.

Authors:  J Eberhard; K Bode; J Hedderich; S Jepsen
Journal:  Clin Oral Investig       Date:  2008-05-24       Impact factor: 3.573

7.  In Vitro Comparison of Diagnostic Accuracy of DIAGNOdent and Digital Radiography for Detection of Secondary Proximal Caries Adjacent to Composite Restorations.

Authors:  Zahra Ghoncheh; Zahra Zonouzy; Nazanin Kiomarsi; Mohammad Javad Kharazifar; Nasim Chiniforush
Journal:  J Lasers Med Sci       Date:  2017-09-27

8.  Evaluation of selective caries removal in deciduous teeth by a fluorescence feedback-controlled Er:YAG laser in vivo.

Authors:  Felix Krause; Andreas Braun; Gabriele Lotz; Susanne Kneist; Søren Jepsen; Jörg Eberhard
Journal:  Clin Oral Investig       Date:  2008-01-26       Impact factor: 3.573

9.  Caries diagnosis using light fluorescence devices: VistaProof and DIAGNOdent.

Authors:  E Betrisey; N Rizcalla; I Krejci; S Ardu
Journal:  Odontology       Date:  2013-03-07       Impact factor: 2.634

10.  A rapid nondestructive method for root dentin moisture measurements: in vitro pilot study.

Authors:  Takashi Komabayashi; Qiang Zhu; Jin Jiang; Kamran E Safavi; Larz S W Spångberg
Journal:  Oral Surg Oral Med Oral Pathol Oral Radiol Endod       Date:  2009-01-20
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.