B Popov1, G Sierra, V Telfort, R Agarwal, V Lanzo.
Abstract
The objective was to test the robustness of an acoustic method to estimate respiratory rates (RR) during treadmill test. The accuracy was assessed by the comparison with simultaneous estimates from a capnograph, using as a common reference a pneumotachometer. Eight subjects without any pulmonary disease were enrolled. Tracheal sounds were acquired using a contact piezoelectric sensor placed on the subject's throat and analyzed using a combined investigation of the sound envelope and frequency content. The capnograph and pneumotachometer were coupled to a face mask worn by the subjects. There was a strong linear correlation between all three methods (r<sup>2</sup>ranged from 0.8 to 0.87), and the SEE ranged from 1.97 to 2.36. As a conclusion, the accuracy of the respiratory rate estimated from tracheal sounds on adult subjects during treadmill stress test was comparable to the accuracy of a commercial capnograph. The heart rate (HR) estimates can also be derived from carotid pulse using the same single sensor placed on the subject's throat. Compared to the pulse oximeter the results show an agreement of acoustic method with r<sup>2</sup>=0.76 and SEE = 3.51.Entities:
Year: 2005 PMID: 17281599 DOI: 10.1109/IEMBS.2005.1615829
Source DB: PubMed Journal: Conf Proc IEEE Eng Med Biol Soc ISSN: 1557-170X