Estimation of intraocular pressure in rabbits with commonly used tonometers.

Ana C Acosta1, Edgar M Espana, Izuru Nose, Marcia Orozco, Hideo Yamamoto, Anna S Venkatraman, Francisco Fantes, Jean-Marie Parel.   

Abstract

BACKGROUND AND OBJECTIVES: To validate accuracy and reproducibility of the Perkins tonometer, pneumatonometer, and Tono-Pen XL (Medtronic Solan, Jacksonville, FL) in estimating intraocular pressure (IOP) in rabbits.
MATERIALS AND METHODS: IOP was increased from 5 to 50 mm Hg in 5-mm increments. Measurements were compared to readings of two digital manometers simultaneously measuring real IOP in the anterior chamber and vitreous cavity. Interobserver accuracy was evaluated using 4 eyes with the Perkins tonometer.
RESULTS: The Perkins tonometer and Tono-Pen XL underestimated IOP and were more accurate at pressures less than 30 mm Hg. No statistically significant difference was found between real IOP and Tono-Pen XL readings. The pneumatonometer overestimated pressures in the low ranges but was accurate at pressures greater than 40 mm Hg. The Tono-Pen XL had more variability than the Perkins tonometer and pneumatonometer at high IOP.
CONCLUSIONS: None of the tonometers are accurate or reproducible in estimating IOP in rabbits over the tested range. Pneumatonometry, although not very accurate, has the advantage of having acceptable variability.

Entities:  

Mesh:

Year:  2007        PMID: 17278535     DOI: 10.3928/15428877-20070101-06

Source DB:  PubMed          Journal:  Ophthalmic Surg Lasers Imaging        ISSN: 1542-8877


  3 in total

1.  Repeatability, reproducibility and agreement of intraocular pressure measurement in rabbits by the TonoVet and Tono-Pen.

Authors:  Di Ma; Chong-Bo Chen; Jiajian Liang; Zhihao Lu; Haoyu Chen; Mingzhi Zhang
Journal:  Sci Rep       Date:  2016-10-12       Impact factor: 4.379

2.  Evaluation of monkey intraocular pressure by rebound tonometer.

Authors:  Wenhan Yu; Guiqun Cao; Jinghua Qiu; Xuyang Liu; Jia Ma; Ni Li; Man Yu; Naihong Yan; Lei Chen; Iok-Hou Pang
Journal:  Mol Vis       Date:  2009-10-27       Impact factor: 2.367

3.  Tono-Pen XL tonometry during application of a suction ring in rabbits.

Authors:  Spyridon K Charisis; Harilaos S Ginis; Georgios A Kounis; Miltiadis K Tsilimbaris
Journal:  BMC Ophthalmol       Date:  2008-07-30       Impact factor: 2.209

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.