[Comparative study of the biometric measurements made by immersion and contact techniques].

Airton Leite Kronbauer1, Fernando Leite Kronbauer, Cláudia Leite Kronbauer.   

Abstract

PURPOSE: To compare clinical biometric findings between measurements of immersion technique and contact technique.
METHODS: Axial length was measured by A-scan in 120 medical examinations in 60 patients with cataract using a non-contact (immersion) and a contact technique in paired-samples by ultrasound.
RESULTS: The mean axial length was found to be 23.19 mm (SD 1.32) with the immersion technique and 22.93 mm (SD 1.32) with the contact technique, using the same transducer probe. The difference of 0.255 mm (SD 0.3) was significant at the 0.01 level. The difference should be taken into account when evaluating the accuracy of IOL calculation. The mean standard deviation between recurrent measures in same eye was found to be 0.04 with the immersion technique and 0.19 with the contact technique. The difference of 0.15 was significant at the 0.01 level. The difference should be taken into account when evaluating the accuracy of reproductivity of technique examination.
CONCLUSION: These data provide benchmark information that can be used to monitor clinical practice and to perform others studies.

Entities:  

Mesh:

Year:  2006        PMID: 17273683     DOI: 10.1590/s0004-27492006000600017

Source DB:  PubMed          Journal:  Arq Bras Oftalmol        ISSN: 0004-2749            Impact factor:   0.872


  1 in total

1.  Comparison of ocular biometry measurements by applanation and immersion A-scan techniques.

Authors:  Dupe S Ademola-Popoola; Donald A Nzeh; Sadiat E Saka; Lateefat B Olokoba; Tokunbo S Obajolowo
Journal:  J Curr Ophthalmol       Date:  2016-02-09
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.