Literature DB >> 17272140

Performance of a radio link between a base station and a medical implant utilising the MICS standard.

Anders J Johansson1.   

Abstract

Modern medical implants are of increasing complexity and with that, the need for fast and flexible communication with them grows. A wireless system is preferable and an inductive link is the most commonly used. But it has the drawback of a very short range, essentially limited to having the external transceiver touching the patient. The Medical Implant Communication System, MICS, is a standard aimed at improving the communication distance. It operates at a higher frequency band between 402 MHz and 405 MHz. We have by simulations and measurements investigated the channel properties of this band and calculated the link performance for a typical implant application. The result is a link speed between a base station and a bedridden patient of 600 kbit bits per second with a bit error rate of 2% in the downlink to the implant and 1 % in the uplink to the base station. Conclusions on the necessary complexity of the base station are also given.

Entities:  

Year:  2004        PMID: 17272140     DOI: 10.1109/IEMBS.2004.1403620

Source DB:  PubMed          Journal:  Conf Proc IEEE Eng Med Biol Soc        ISSN: 1557-170X


  1 in total

1.  Enhancing the Performance of Medical Implant Communication Systems through Cooperative Diversity.

Authors:  Barnabás Hegyi; János Levendovszky
Journal:  Int J Telemed Appl       Date:  2010-03-30
  1 in total