Reduced best spectacle-corrected visual acuity from inserting a thicker Intacs above and thinner Intacs below in keratoconus.

Colin C K Chan1, Brian S Boxer Wachler.   

Abstract

PURPOSE: To report a case of decreased best spectacle-corrected visual acuity (BSCVA) 2 months after Intacs implantation.
METHODS: A 33-year-old woman with keratoconus and contact lens intolerance underwent Intacs surgery in the left eye at another institution. Two segments were used--a thinner one (0.25 mm) below the cone and a thicker one (0.35 mm) above the cone.
RESULTS: Two months postoperatively, the patient presented to our practice with BSCVA reduced from 20/20 to 20/30. The superior Intacs segment was explanted, the inferior segment was exchanged for a thicker one (0.35 mm), and collagen cross-linking with riboflavin treatment was performed. This resulted in visual, topographic, and refractive improvement with BSCVA returning to 20/20.
CONCLUSIONS: Single inferior segment Intacs may be more appropriate for paracentral and peripheral cones. Collagen cross-linking may help cause further flattening. Using asymmetrical segments, with the thicker segment above the cone, may increase distortions and result in loss of BSCVA.

Entities:  

Mesh:

Year:  2007        PMID: 17269250     DOI: 10.3928/1081-597X-20070101-15

Source DB:  PubMed          Journal:  J Refract Surg        ISSN: 1081-597X            Impact factor:   3.573


  2 in total

1.  Bevel toric multicurve rigid gas-permeable lens for keratoconus.

Authors:  Ryoji Yanai; Kiichi Ueda; Koh-Hei Sonoda
Journal:  Jpn J Ophthalmol       Date:  2012-12-07       Impact factor: 2.447

2.  Intrastromal corneal ring implants for corneal thinning disorders: an evidence-based analysis.

Authors: 
Journal:  Ont Health Technol Assess Ser       Date:  2009-04-01
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.