Scanning electron microscopic analysis of defects in polymer coatings of three commercially available stents: comparison of BiodivYsio, Taxus and Cypher stents.

Yoritaka Otsuka1, Nicolas A F Chronos, Robert P Apkarian, Keith A Robinson.   

Abstract

BACKGROUND: Although the use of polymer-based drug-eluting stents appears to markedly reduce the risk of in-stent restenosis, there are concerns about their safety including polymer layer integrity.
OBJECTIVES: The objective of this study was to investigate the morphology of the polymer layer of 3 commercially available polymercoated stents, including the effects of balloon catheter expansion, by scanning electron microscopy (SEM).
METHODS: We assessed discontinuities and other irregularities in the polymer layer of BiodivYsio, Taxus and Cypher stents by SEM after balloon expansion in saline solution at 37 degrees C.
RESULTS: Distinctive polymer layer morphologies were found among the 3 stent types, including responses to balloon expansion and withdrawal. The BiodivYsio stent showed no waving or other irregularities on the outer surface, but excess polymer was present on stent edges and polymer was peeled off from the inner surface. The Taxus stent showed no irregularities on the outer surface, but there were polymer bridgings across strut loops and linear cracking of the bridges, as well as inner surface polymer defects with bare-metal exposure. The Cypher stent showed a rough surface with irregularities and waving on the outer surface. There also appeared to be polymer defects with bare-metal exposure in the loop region and peeling of the top-coated polymer layer in the loop.
CONCLUSIONS: We found several types of defects in the polymer layers on commercially available polymer-coated stents. Some of these indicate potential risks of thrombosis, coronary microembolism of polymer layer pieces and late inflammatory or neointimal reactions.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 17268041

Source DB:  PubMed          Journal:  J Invasive Cardiol        ISSN: 1042-3931            Impact factor:   2.022


  15 in total

Review 1.  Retractable-needle catheters: an update on local drug delivery in coronary interventions.

Authors:  Paolo Angelini; Wijay Bandula
Journal:  Tex Heart Inst J       Date:  2008

2.  Scanning electron microscopic analysis of polymer damage in drug-eluting stents during multiple stenting.

Authors:  Aiko Shimokado; Takashi Kubo; Hirotoshi Utsunomiya; Hisashi Yamasaki; Takashi Tanimoto; Yasushi Ino; Yong Liu; Ikuko Teraguchi; Yuichi Ozaki; Yasutsugu Shiono; Makoto Orii; Kunihiro Shimamura; Takashi Yamano; Tomoyuki Yamaguchi; Kumiko Hirata; Toshio Imanishi; Takashi Akasaka
Journal:  Int J Cardiovasc Imaging       Date:  2013-09-13       Impact factor: 2.357

3.  Tailoring biomaterial surface properties to modulate host-implant interactions: implication in cardiovascular and bone therapy.

Authors:  Settimio Pacelli; Vijayan Manoharan; Anna Desalvo; Nikita Lomis; Kartikeya Singh Jodha; Satya Prakash; Arghya Paul
Journal:  J Mater Chem B       Date:  2015-10-16       Impact factor: 6.331

4.  Endothelial retention and phenotype on carbonized cardiovascular implant surfaces.

Authors:  Christopher M Frendl; Scott M Tucker; Nadeem A Khan; Mandy B Esch; Shrinidhi Kanduru; Thong M Cao; Andrés J García; Michael R King; Jonathan T Butcher
Journal:  Biomaterials       Date:  2014-06-20       Impact factor: 12.479

5.  Stability and bioactivity of pepCD47 attachment on stainless steel surfaces.

Authors:  Vaishali V Inamdar; Emmett Fitzpatrick; Ivan Alferiev; Chandrasekaran Nagaswami; Lynn A Spruce; Hossein Fazelinia; George Bratinov; Steven H Seeholzer; Robert J Levy; Ilia Fishbein; Stanley J Stachelek
Journal:  Acta Biomater       Date:  2020-01-11       Impact factor: 8.947

6.  Delivery of paclitaxel from cobalt-chromium alloy surfaces without polymeric carriers.

Authors:  Gopinath Mani; Celia E Macias; Marc D Feldman; Denes Marton; Sunho Oh; C Mauli Agrawal
Journal:  Biomaterials       Date:  2010-04-15       Impact factor: 12.479

7.  Changes to polymer surface of drug-eluting stents during balloon expansion.

Authors:  Scott J Denardo; Paul L Carpinone; David M Vock; Christopher D Batich; Carl J Pepine
Journal:  JAMA       Date:  2012-05-23       Impact factor: 56.272

8.  Evaluation of the effect of expansion and shear stress on a self-assembled endothelium mimicking nanomatrix coating for drug eluting stents in vitro and in vivo.

Authors:  Adinarayana Andukuri; IlJae Min; Patrick Hwang; Grant Alexander; Lauren E Marshall; Joel L Berry; Timothy M Wick; Yoon Ki Joung; Young-Sup Yoon; Brigitta C Brott; Dong Keun Han; Ho-Wook Jun
Journal:  Biofabrication       Date:  2014-07-22       Impact factor: 9.954

9.  Polymer Coating Integrity, Thrombogenicity and Computational Fluid Dynamics Analysis of Provisional Stenting Technique in the Left Main Bifurcation Setting: Insights from an In-Vitro Model.

Authors:  Marek Milewski; Chen Koon Jaryl Ng; Pawel Gąsior; Shaoliang Shawn Lian; Su Xiao Qian; Shengjie Lu; Nicolas Foin; Elvin Kedhi; Wojciech Wojakowski; Hui Ying Ang
Journal:  Polymers (Basel)       Date:  2022-04-22       Impact factor: 4.967

10.  Characterization of Porous TiO2 Surfaces Formed on 316L Stainless Steel by Plasma Electrolytic Oxidation for Stent Applications.

Authors:  Zhiguang Huan; Lidy E Fratila-Apachitei; Iulian Apachitei; Jurek Duszczyk
Journal:  J Funct Biomater       Date:  2012-05-11
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.