Flow rate, syringe size and architecture are critical to start-up performance of syringe pumps.

S B Neff1, T A Neff, S Gerber, M M Weiss.   

Abstract

BACKGROUND AND
OBJECTIVE: Significant start-up delays are inherent to syringe infusion pumps, particularly at low infusion rates, as routinely used in children's anaesthesia and intensive care. Such delays are mainly the result of engagement of gears in the mechanical drive or compliance of the syringe assembly. The purpose of the present study was to determine the effect of flow rate, syringe size and syringe architecture on fluid delivery during infusion start-up.
METHODS: Elapsed time from infusion start to achievement of steady-state flow was gravimetrically determined for various infusion rates (0.1, 0.5, 1 mL h-1), different syringe sizes (10-, 20-, 30-, 50-mL) and syringes of two different brands (BD and Codan). Four measurements for each condition were performed with two identical Alaris Asena GH syringe infusion pumps (total of eight experiments). Statistical analysis was done by two-way ANOVA with Bonferroni's post-test; P < 0.05 was considered significant.
RESULTS: Start-up time was from 3.6 +/- 0.9 min (BD 10-mL syringe, 1.0 mL h-1) to 74.5 +/- 26.6 min (BD 50-mL syringe, 0.1 mL h-1). Overall, the start-up time markedly increased with lower flow rate (0.1 mL h-1 vs. 1 mL h-1; P < 0.0001), larger syringe size (50 mL vs. 10 mL; P < 0.01), and the BD brand in comparison with the Codan syringes (P < 0.01).
CONCLUSIONS: Highest possible flow rate, smaller sized syringes and syringe plungers with reduced compressibility should be preferred in order to avoid significant start-up delays in fluid delivery.

Entities:  

Mesh:

Year:  2007        PMID: 17261217     DOI: 10.1017/S0265021506002328

Source DB:  PubMed          Journal:  Eur J Anaesthesiol        ISSN: 0265-0215            Impact factor:   4.330


  5 in total

1.  Performance of the BBraun perfusor space syringe driver under hyperbaric conditions.

Authors:  Lachlan Frawley; Bridget Devaney; Theo Tsouras; Geoff Frawley
Journal:  Diving Hyperb Med       Date:  2017-03       Impact factor: 0.887

2.  Capacitive coupling synchronizes autonomous microfluidic oscillators.

Authors:  Sasha Cai Lesher-Pérez; Chao Zhang; Shuichi Takayama
Journal:  Electrophoresis       Date:  2018-02-23       Impact factor: 3.535

3.  Delay and stability of central venous administration of norepinephrine in children: a bench study.

Authors:  Mehdi Oualha; Mikaël Capelo; Odile Spreux-Varoquaux; Isabelle Drouet-Chaillou; Jean-Marc Tréluyer; Philippe Hubert; Fabrice Lesage
Journal:  Eur J Pediatr       Date:  2013-08-11       Impact factor: 3.183

4.  Accuracy assessment of a PION TCI pump based on international standards.

Authors:  Il Doh; Seok Hwan Lee; Yong-Hun Lee; Bokyoung Jeon; Byung-Moon Choi; Gyu-Jeong Noh
Journal:  Anesth Pain Med (Seoul)       Date:  2019-10-31

5.  Development of a Population Pharmacokinetic Model of Busulfan in Children and Evaluation of Different Sampling Schedules for Precision Dosing.

Authors:  Efthymios Neroutsos; Ricardo Nalda-Molina; Anna Paisiou; Kalliopi Zisaki; Evgenios Goussetis; Alexandros Spyridonidis; Vasiliki Kitra; Stelios Grafakos; Georgia Valsami; Aristides Dokoumetzidis
Journal:  Pharmaceutics       Date:  2022-03-15       Impact factor: 6.321

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.