Laura Pisani1, Roland Bammer, Gary Glover.
Abstract
A restricted field of view (rFOV) approach for imaging a dynamic time series of volumes of limited spatial extent within a larger subject is described. The shorter readout with rFOV-MRI can be exploited to either limit image artifacts or increase spatial resolution. To accomplish rFOV imaging of a multislice volume for a dynamic series, an outer volume suppression (OVS) preparation that saturates signal external to a cylinder through the subject is followed by slice-selective excitation and a spiral readout. The pass- and stopband efficiencies of the OVS in an agar gel phantom were 97% (+/-1.5%) and 3% (+/-1%), respectively. Profiles of the temporal signal-to-noise ratio (SNR) were measured in a phantom and an adult brain. The rFOV sequence reduced distortions from off-resonance signal and T2*-induced blurring compared to a conventional sequence. Sequence utility is demonstrated for high-resolution rFOV functional MRI (fMRI) in the visual cortex. The rFOV sequence may prove to be useful for other multislice dynamic and high-resolution imaging applications. Copyright (c) 2007 Wiley-Liss, Inc.Entities:
Mesh:
Substances:
Year: 2007 PMID: 17260360 DOI: 10.1002/mrm.21115
Source DB: PubMed Journal: Magn Reson Med ISSN: 0740-3194 Impact factor: 4.668