Interchangeability of measurements of CA 19-9 in serum with four frequently used assays: an update.

Rita Passerini1, Daniela Riggio, Michela Salvatici, Laura Zorzino, Davide Radice, Maria Teresa Sandri.   

Abstract

BACKGROUND: CA 19-9 is a marker principally related to pancreatic and gall bladder cancer. Although its determination has no value in screening for these malignancies, it is used in post-operative monitoring and during chemotherapeutic treatment of confirmed disease. Measurements during follow-up must be comparable and must be performed with standard, validated methods.
METHODS: We compared four routinely used analytical systems for CA 19-9 determination: the Architect i2000 and AxSYM systems from Abbott Laboratories, the Elecsys 1010 from Roche Diagnostics, and the KRYPTOR system from Brahms Diagnostics. We evaluated the analytical performance of the four systems and compared measurements of CA 19-9 values, which covered the whole analytical range.
RESULTS: The analytical performance and accuracy of the four systems were fairly good, but Passing-Bablok regression and mountain plots showed significant differences in CA 19-9 values measured with the four platforms.
CONCLUSIONS: Our data indicate that during tumor follow-up, the use of the same system is appropriate to avoid the risk of a variation due to the method rather than the disease. Moreover, whenever a change in analytical equipment is required, careful analysis of CA 19-9 results must be undertaken.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 17243924     DOI: 10.1515/CCLM.2007.003

Source DB:  PubMed          Journal:  Clin Chem Lab Med        ISSN: 1434-6621            Impact factor:   3.694


  3 in total

1.  Comparison of LUMIPULSE(®) G1200 With Kryptor and Modular E170 for the Measurement of Seven Tumor Markers.

Authors:  Julien Marlet; Maguy Bernard
Journal:  J Clin Lab Anal       Date:  2014-10-03       Impact factor: 2.352

2.  Definitive Characterization of CA 19-9 in Resectable Pancreatic Cancer Using a Reference Set of Serum and Plasma Specimens.

Authors:  Brian B Haab; Ying Huang; Seetharaman Balasenthil; Katie Partyka; Huiyuan Tang; Michelle Anderson; Peter Allen; Aaron Sasson; Herbert Zeh; Karen Kaul; Doron Kletter; Shaokui Ge; Marshall Bern; Richard Kwon; Ivan Blasutig; Sudhir Srivastava; Marsha L Frazier; Subrata Sen; Michael A Hollingsworth; Jo Ann Rinaudo; Ann M Killary; Randall E Brand
Journal:  PLoS One       Date:  2015-10-02       Impact factor: 3.240

3.  Immunosensor for Pancreatic Cancer Based on Electrospun Nanofibers Coated with Carbon Nanotubes or Gold Nanoparticles.

Authors:  Juliana C Soares; Leonardo E O Iwaki; Andrey C Soares; Valquiria C Rodrigues; Matias E Melendez; José Humberto T G Fregnani; Rui M Reis; Andre L Carvalho; Daniel S Corrêa; Osvaldo N Oliveira
Journal:  ACS Omega       Date:  2017-10-19
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.