Closed loop stimulation improves ejection fraction in pediatric patients with pacemaker and ventricular dysfunction.

Fabrizio Drago1, Massimo S Silvetti, Antonella De Santis, Giovanni Fazio, Gianluca Biancalana, Giorgia Grutter, Gabriele Rinelli.   

Abstract

BACKGROUND: The aim of this prospective study was to evaluate the effect of the closed loop stimulation (CLS) on the ejection fraction in pediatric patients, affected by complete atrioventricular block (CAVB) or CAVB and sinus node dysfunction (SND), with a previously implanted pacemaker (PM) and ventricular dysfunction. The role of electrical therapy in the treatment of pediatric patients with congenital atrioventricular (AV) blocks has been shown. Conventional right ventricular pacing seems to affect ventricular function. Up to now, the feasibility and the long-term results of biventricular pacing in pediatric patients were not entirely clear.
METHODS: In eight pediatric patients with a previously implanted single or dual chamber PM, ventricular dysfunction, and CAVB or SND and CAVB, a dual chamber PM INOS(2+)-CLS (Biotronik GmbH, Berlin, Germany) was implanted. The effect of the physiological modulation of CLS pacing mode on the ejection fraction was evaluated by Echo-Doppler examination. Measurements were performed before the substitution of the old PM and for up to 2 years of follow-up.
RESULTS: All patients showed correct electrical parameters at implantation and during follow-up. The mean value of the ejection fraction measured before the replacement of the old PM was 36 +/- 7%, while after 2 years it was 47 +/- 1% (P < 0.003). No patient showed any worsening of the ejection fraction, while only one showed no improvement.
CONCLUSIONS: DDD-CLS pacing seems to improve ventricular function in pediatric patients with CAVB and/SND in spite of the use of the apical right conventional stimulation.

Entities:  

Mesh:

Year:  2007        PMID: 17241312     DOI: 10.1111/j.1540-8159.2007.00576.x

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  2 in total

1.  Smart sensors for maintaining physiologic homeostasis.

Authors:  David C Klonoff
Journal:  J Diabetes Sci Technol       Date:  2011-05-01

2.  Atrial auto-short phenomenon as a rare cause of ventricular lead failure in a pediatric dual chamber pacemaker patient.

Authors:  Nathalie Noessler; Martin Koestenberger; Stefan Kurath-Koller
Journal:  Pacing Clin Electrophysiol       Date:  2020-02-26       Impact factor: 1.976

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.