Performance characteristics of ELISAs for monitoring ketamine exposure.

Mei-Han Huang1, Meng-Yen Wu, Chih-Hung Wu, Jin-Lian Tsai, Hei-Hwa Lee, Ray H Liu.   

Abstract

BACKGROUND: Growing misuse of ketamine necessitates the development of high throughput testing approaches. Two commercial enzyme-linked immunosorbent assays (ELISA) for ketamine have recently become available and were adapted for this development.
METHOD: The newly available ketamine ELISA reagents were examined to better understand their cross-reacting, calibration and other performance characteristics. ELISA apparent analyte concentrations were also correlated against ketamine concentrations as determined by GC-MS to examine the relationship between these 2 parameters. RESULT: Both adapting ketamine as the targeted analyte, reagent from International Diagnostic Systems (IDS) also responded very significantly to the metabolites of ketamine (norketamine and dehydronorketamine), while the NEOGEN reagent responded very specifically to ketamine.
CONCLUSION: NEOGEN ELISA test data exhibit better correlation with the ketamine concentration as determined by GC-MS. It can be more reliably used as the preliminary test method in the 2-step approach now routinely adapted in workplace drug testing programs. Using 100 ng/ml ketamine as the GC-MS cutoff, the corresponding ELISA cutoff value is approximately 110-120 ng/ml. With significantly higher responses to ketamine metabolites, IDS reagent can detect specimens with much lower ketamine/metabolites concentrations and can better meet other testing requirements.

Entities:  

Mesh:

Substances:

Year:  2006        PMID: 17239837     DOI: 10.1016/j.cca.2006.12.013

Source DB:  PubMed          Journal:  Clin Chim Acta        ISSN: 0009-8981            Impact factor:   3.786


  3 in total

1.  Rapid detection and quantitation of drugs-of-abuse by wooden-tip electrospray ionization mass spectrometry.

Authors:  Tsz-Tsun Ng; Pui-Kin So; Bin Hu; Zhong-Ping Yao
Journal:  J Food Drug Anal       Date:  2018-10-01       Impact factor: 6.157

2.  A label-free immunosensor for ultrasensitive detection of ketamine based on quartz crystal microbalance.

Authors:  Ya Yang; Yifeng Tu; Xiaoshu Wang; Jinyin Pan; Yun Ding
Journal:  Sensors (Basel)       Date:  2015-04-13       Impact factor: 3.576

3.  Association of urinary ketamine and APOA1 levels with bladder dysfunction in ketamine abusers revealed via proteomics and targeted metabolite analyses.

Authors:  Jo-Chuan Liu; Yi-Ting Chen; Ya-Ju Hsieh; Chia-Chun Wu; Ming-Chyi Huang; Yu-Chao Hsu; Chun-Te Wu; Chih-Ken Chen; Srinivas Dash; Jau-Song Yu
Journal:  Sci Rep       Date:  2021-05-05       Impact factor: 4.379

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.