Literature DB >> 17239609

Safety of a thermal vessel sealer on main pulmonary vessels.

Tunc Lacin1, Hasan Fevzi Batirel, Kursat Ozer, Alper Demirutku, Rengin Ahiskali, Mustafa Yuksel.   

Abstract

OBJECTIVE: LigaSure (Valleylab, Tyco Healthcare, Boulder, CO) is a novel instrument with widespread use in various surgical disciplines for vessel sealing. It uses heat energy to denature collagen and elastin and achieve vessel sealing. We investigated the safety of LigaSure in pulmonary arteries (PA) and veins (PV).
METHODS: Twelve sheep were anesthetized and endotracheally intubated. Six underwent right lower lobectomy (Group 1) and six right upper lobectomy (Group 2). All of the vessels in both groups were divided with LigaSure. Diameters of vessels were measured. Intraoperatively, pulmonary artery pressures (PAP) were monitored in Group 1. Following vascular division, ephedrine was injected to increase PAP in Group 1. In Group 2, animals were followed and euthanized at 7 days. Both immediate and 7 day samples were obtained from divided vessel stumps. Conventional histology was done. Statistical analysis was performed using t-test and Fisher's exact test.
RESULTS: Mean diameter of PAs (n=9) divided in Group 1 was 8.3mm (3-11) and of PVs (n=8) was 10.4mm (4-15). Mean PAP increased from 18 (27/9) to 27 (45/18)mmHg after ephedrine injection. Dehiscence rate was 2/6 in PAs and 3/6 in PVs larger than 9 mm in diameter. Mean diameter of PAs (n=8) divided in Group 2 was 5.7 mm (3-7) and of PVs (n=9) was 4.6mm (4-6). No early or late (7 days) vascular disruption was seen in Group 2. Bleeding incidence was significantly different between groups (Fisher's exact test p=0.015). Overall no dehiscence was observed in vessels less than 7 mm in diameter. Histology of intraoperative samples showed thermal injury and necrosis. Seven day samples showed short segment necrosis limited to the vessel wall, fibrosis, thrombus formation without inflammation, and minimal granulation tissue.
CONCLUSIONS: LigaSure achieves perfect sealing in pulmonary vessels less than 7 mm in diameter in sheep intra- and postoperatively even following a pressure challenge. It can safely be used in segmental branches of PAs and PVs during open or thoracoscopic surgery.

Entities:  

Mesh:

Year:  2007        PMID: 17239609     DOI: 10.1016/j.ejcts.2006.11.038

Source DB:  PubMed          Journal:  Eur J Cardiothorac Surg        ISSN: 1010-7940            Impact factor:   4.191


  2 in total

1.  Postoperative efficacy and safety of vessel sealing: an experimental study on carotid arteries of the pig.

Authors:  Stéphane V Berdah; Christiaan Hoff; Peiman Hossein Poornoroozy; Peter Razek; Yves Van Nieuwenhove
Journal:  Surg Endosc       Date:  2012-02-21       Impact factor: 4.584

2.  Intralobar sequestration associated with cystic adenomatoid malformation: diagnostic and thoracoscopic pitfalls.

Authors:  Smart Zeidan; Geraldine Hery; Ferderic Lacroix; Guillaume Gorincour; Alain Potier; Jean Christophe Dubus; Jean-Michel Guys; Pascal de Lagausie
Journal:  Surg Endosc       Date:  2008-10-15       Impact factor: 4.584