Are ankle and toe brachial indices (ABI-TBI) obtained by a pocket Doppler interchangeable with those obtained by standard laboratory equipment?

Phyllis A Bonham1, Marguerite Cappuccio, Tara Hulsey, Yvonne Michel, Teresa Kelechi, Carolyn Jenkins, Jacob Robison.   

Abstract

PURPOSE: The aim of this study was to determine if the ankle and toe brachial indices obtained by an experienced registered nurse (RN) using a pocket Doppler were within acceptable levels of agreement with those obtained by a registered vascular technologist (RVT) using standard laboratory equipment.
DESIGN: A within-subjects comparative design was used. SETTING AND
SUBJECTS: Thirty subjects who were referred to a vascular laboratory for arterial studies were recruited. All tests were performed in the outpatient vascular laboratory of a large, urban medical center.
METHODOLOGY: Ankle and toe brachial indices were measured on each subject by the RN and the RVT during each visit. Data were analyzed using the Bland-Altman method to assess the level of agreement between the RN's pocket Doppler and the RVT's standard laboratory equipment.
RESULTS: Differences between each instrument's ankle brachial indices were within the a priori 15% limit of agreement. Differences between each instrument's toe brachial indices exceeded the a priori 15% limit of agreement.
CONCLUSION: The ankle brachial index obtained by using a pocket Doppler by an experienced RN is interchangeable with vascular laboratory testing for detection of lower extremity arterial disease (LEAD). The pocket Doppler-derived toe brachial index was not interchangeable with vascular laboratory testing for detection of LEAD.

Entities:  

Mesh:

Year:  2007        PMID: 17228206     DOI: 10.1097/00152192-200701000-00007

Source DB:  PubMed          Journal:  J Wound Ostomy Continence Nurs        ISSN: 1071-5754            Impact factor:   1.741


  5 in total

1.  An evidence-based diagnostic classification system for low back pain.

Authors:  Robert Vining; Eric Potocki; Michael Seidman; A Paige Morgenthal
Journal:  J Can Chiropr Assoc       Date:  2013-09

2.  Interrater and intrarater reliability of photoplethysmography for measuring toe blood pressure and toe-brachial index in people with diabetes mellitus.

Authors:  Christopher Scanlon; Kris Park; David Mapletoft; Lindy Begg; Joshua Burns
Journal:  J Foot Ankle Res       Date:  2012-06-07       Impact factor: 2.303

3.  The reliability of toe systolic pressure and the toe brachial index in patients with diabetes.

Authors:  Mary T Romanos; Anita Raspovic; Byron M Perrin
Journal:  J Foot Ankle Res       Date:  2010-12-22       Impact factor: 2.303

4.  Pocket Doppler and vascular laboratory equipment yield comparable results for ankle brachial index measurement.

Authors:  Saskia P A Nicolaï; Lotte M Kruidenier; Ellen V Rouwet; Liliane Wetzels-Gulpers; Constantijn A M Rozeman; Martin H Prins; Joep A W Teijink
Journal:  BMC Cardiovasc Disord       Date:  2008-10-07       Impact factor: 2.298

5.  The reliability of the ankle brachial index: a systematic review.

Authors:  Sarah Casey; Sean Lanting; Christopher Oldmeadow; Vivienne Chuter
Journal:  J Foot Ankle Res       Date:  2019-08-02       Impact factor: 2.303

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.