Method validation for the analysis of 169 pesticides in soya grain, without clean up, by liquid chromatography-tandem mass spectrometry using positive and negative electrospray ionization.

Ionara R Pizzutti1, André de Kok, Renato Zanella, Martha B Adaime, Maurice Hiemstra, Cristine Wickert, Osmar D Prestes.   

Abstract

Part of a comprehensive study on the comparison of different extraction methods, GC-MS(/MS) and LC-MS/MS detection methods and modes, for the analysis of soya samples is described in this paper. The validation of an acetone-based extraction method for analysis of 169 pesticides in soya, using LC-MS/MS positive and negative electrospray ionisation (ESI) mode, is reported. Samples (5 g) were soaked with 10 g water and subsequently extracted with 100 mL of a mixture of acetone, dichloromethane and light petroleum (1:1:1), in the presence of 15 g anhydrous sodium sulphate. After centrifugation, aliquots of the extract were evaporated and reconstituted in 1.0 mL of methanol, before direct injection of the final extract (corresponding with 0.05 g soya mL(-1)) into the LC-MS/MS system. Linearity, r(2) of calibration curves, instrument limit of detection/quantitation (LOD/LOQ) and matrix effect were evaluated, based on seven concentrations measured in 6-fold. Good linearity (at least r(2)> or =0.99) of the calibration curves was obtained over the range from 0.1 or 0.25 to 10.0 ng mL(-1), corresponding with pesticide concentrations in soya bean extract of 2 or 5-200 microg kg(-1). Instrument LOD values generally were 0.1 or 0.25 ng mL(-1). Matrix effects were negligible for approximately 90% of the pesticides. The accuracy, precision and method LOQ were determined via recovery experiments, spiking soya at 10, 50, 100 microg kg(-1), six replicates per level. In both ESI modes, method LOQ values were mostly 10 or 50 microg kg(-1) and more than 70% of pesticides analysed by each mode met the acceptability criteria of recovery (70-120%) and RSD (< or =20%), at one or more of the three levels studied. A fast, easy and efficient method with acceptable performance was achieved for a difficult matrix as soya, without cleanup.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 17222861     DOI: 10.1016/j.chroma.2006.12.030

Source DB:  PubMed          Journal:  J Chromatogr A        ISSN: 0021-9673            Impact factor:   4.759


  2 in total

1.  Method development and validation of ten pyrethroid insecticides in edible mushrooms by Modified QuEChERS and gas chromatography-tandem mass spectrometry.

Authors:  Fajun Tian; Chengkui Qiao; Jing Luo; Linlin Guo; Tao Pang; Rongli Pang; Jun Li; Caixia Wang; Ruiping Wang; Hanzhong Xie
Journal:  Sci Rep       Date:  2020-04-27       Impact factor: 4.379

2.  Modification and re-validation of the ethyl acetate-based multi-residue method for pesticides in produce.

Authors:  Hans G J Mol; Astrid Rooseboom; Ruud van Dam; Marleen Roding; Karin Arondeus; Suryati Sunarto
Journal:  Anal Bioanal Chem       Date:  2007-06-12       Impact factor: 4.142

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.