Coronary artery stenosis quantification using multislice computed tomography.

Marc Dewey1, Wolfgang Rutsch, Dirk Schnapauff, Florian Teige, Bernd Hamm.   

Abstract

RATIONALE AND
OBJECTIVES: Reliable noninvasive detection of stenoses with multislice computed tomography (MSCT) is feasible. This study's aim was to analyze the agreement, correlation, and reliability of MSCT with conventional coronary angiography as the reference standard for quantification of coronary artery stenoses.
MATERIALS AND METHODS: A total of 118 significant (at least 50%) coronary artery stenoses with a reference vessel diameter of at least 1.5 mm in 62 patients were analyzed by MSCT using 16 detector rows (Aquilion, Toshiba, Otawara, Japan), multisegment reconstruction, and voxel sizes of 0.35x0.35x0.5 mm. The degree of stenosis on MSCT and quantitative coronary angiography (QCA) was measured by correlating the difference between the reference vessel diameter (average of 2 measurements directly proximal and distal to the stenosis) and the stenotic vessel diameter to the reference vessel diameter.
RESULTS: Correlation between the percent diameter stenosis determined by MSCT (78.2+/-13.6%) and QCA (76.0+/-14.8%) was significant (P<0.001) but only moderately so (R=0.51). Bland-Altman analysis revealed no systematic under- or overestimation with MSCT but large limits of agreements (+/-27.6%). Also the limits of agreement for interobserver agreement (reliability) of MSCT data were considerably large (+/-24.8%). Among the 27 coronary artery stenoses with a reference diameter of at least 3.5 mm, there was improved correlation (R=0.80) and the limits of agreement between MSCT and QCA were significantly smaller (+/-17.3%, P<0.008). The agreement between MSCT and QCA was not significantly different for stenoses with no calcification or only calcium spots (+/-28.2%) as compared with those with moderate-or-severe calcifications (+/-27.3%; P=0.8). MSCT allowed correct classification of coronary stenoses into low-grade (below 75%) and high-grade stenoses (at least 75%), in 62% (73 of 118).
CONCLUSIONS: The accuracy and reliability of coronary artery stenosis quantification with MSCT using isotropic voxel sizes and multisegment reconstruction is still too low to recommend routine clinical application because of rather low agreement, correlation, and reliability. Despite these limitations, the current results demonstrate the potential of MSCT for reliable and accurate quantification of coronary artery stenoses in the near future provided that further improvements in spatial and temporal resolution will be achieved.

Entities:  

Mesh:

Year:  2007        PMID: 17220725     DOI: 10.1097/01.rli.0000251569.01317.60

Source DB:  PubMed          Journal:  Invest Radiol        ISSN: 0020-9996            Impact factor:   6.016


  10 in total

Review 1.  CT coronary angiography of chronic total occlusions of the coronary arteries: how to recognize and evaluate and usefulness for planning percutaneous coronary interventions.

Authors:  John Hoe
Journal:  Int J Cardiovasc Imaging       Date:  2009-01-23       Impact factor: 2.357

2.  Accuracy of dual-source computed tomography in quantitative assessment of low density coronary stenosis--a motion phantom study.

Authors:  Michael Toepker; Christopher L Schlett; Thomas Irlbeck; Amir A Mahabadi; Fabian Bamberg; Christiane Leidecker; Patrick Donnelly; Udo Hoffmann
Journal:  Eur Radiol       Date:  2009-09-02       Impact factor: 5.315

3.  Four-dimensional computed tomographic imaging in the wrist: proof of feasibility in a cadaveric model.

Authors:  Shian-Chao Tay; Andrew N Primak; Joel G Fletcher; Bernhard Schmidt; Kimberly K Amrami; Richard A Berger; Cynthia H McCollough
Journal:  Skeletal Radiol       Date:  2007-09-06       Impact factor: 2.199

4.  CT comparison of visual and computerised quantification of coronary stenosis according to plaque composition.

Authors:  Doo Kyoung Kang; Nae Jung Im; Soon Mo Park; Hong Seok Lim
Journal:  Eur Radiol       Date:  2010-10-02       Impact factor: 5.315

5.  Comparison between different kernel reformatting filters in 3D quantitative analysis of MDCT coronary angiography.

Authors:  R Malagò; A Pezzato; C Barbiani; W Mantovani; G Caliari; U Alfonsi; D Tavella; R Pozzi Mucelli
Journal:  Radiol Med       Date:  2011-09-02       Impact factor: 3.469

6.  Accuracy of quantitative coronary angiography with computed tomography and its dependency on plaque composition: plaque composition and accuracy of cardiac CT.

Authors:  Lars Husmann; Oliver Gaemperli; Tiziano Schepis; Hans Scheffel; Ines Valenta; Tobias Hoefflinghaus; Paul Stolzmann; Lotus Desbiolles; Bernhard A Herzog; Sebastian Leschka; Borut Marincek; Hatem Alkadhi; Philipp A Kaufmann
Journal:  Int J Cardiovasc Imaging       Date:  2008-06-19       Impact factor: 2.357

7.  Influence of heart rate on diagnostic accuracy and image quality of 16-slice CT coronary angiography: comparison of multisegment and halfscan reconstruction approaches.

Authors:  Marc Dewey; Florian Teige; Michael Laule; Bernd Hamm
Journal:  Eur Radiol       Date:  2007-07-17       Impact factor: 5.315

8.  Quantification of three-dimensional computed tomography angiography for evaluating coronary luminal stenosis using digital subtraction angiography as the standard of reference.

Authors:  Wei Guo; Xin Liu; Zhifan Gao; Sandeep Pirbhulal; Wenhua Huang; Wan-Hua Lin; Heye Zhang; Ning Tan; Yuan-Ting Zhang
Journal:  Biomed Eng Online       Date:  2015-05-30       Impact factor: 2.819

9.  Evaluation of computed tomography in patients with atypical angina or chest pain clinically referred for invasive coronary angiography: randomised controlled trial.

Authors:  Marc Dewey; Matthias Rief; Peter Martus; Benjamin Kendziora; Sarah Feger; Henryk Dreger; Sascha Priem; Fabian Knebel; Marko Böhm; Peter Schlattmann; Bernd Hamm; Eva Schönenberger; Michael Laule; Elke Zimmermann
Journal:  BMJ       Date:  2016-10-24

10.  Is there a gender difference in noninvasive coronary imaging? Multislice computed tomography for noninvasive detection of coronary stenoses.

Authors:  Marc Dewey; Wolfgang Rutsch; Bernd Hamm
Journal:  BMC Cardiovasc Disord       Date:  2008-01-29       Impact factor: 2.298

  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.