Initial clinical experience with matrix detachable coils for the treatment of intracranial aneurysms.

Yuichi Murayama1, Fernando Viñuela, Akira Ishii, Yih-Lin Nien, Ichiro Yuki, Gary Duckwiler, Reza Jahan.   

Abstract

OBJECT: The Matrix detachable coil is a new bioactive, bioabsorbable coil used in the endovascular embolization of intracranial aneurysms. It has a platinum core covered with a bioactive, bioabsorbable polymer (polyglycolic acid/lactide). The authors report on their initial midterm clinical experience with the first-generation Matrix detachable coil.
METHODS: One hundred twelve patients harboring 118 aneurysms were treated using Matrix coils. Forty-nine aneurysms (41.5%) were associated with acute subarachnoid hemorrhage (SAH). Twenty-four lesions (49%) were harbored by patients with Hunt and Hess Grade I, 11 (23.4%) by patients with Grade II, eight (16.3%) by those with Grade III, and six (12.2%) by those with Grade IV. Four aneurysms (3.4%) were harbored by patients who had presented with nonacute SAH. Sixty-five aneurysms (55%) were unruptured. Fifty-seven lesions (48.3%) were small with a small neck, 29 (24.6%) were small with a wide neck, 30 (25.4%) were large, and two (1.7%) were giant. All patients were followed up to obtain angiography and clinical outcome data. Technical complications occurred in six patients: two thromboembolic complications and four aneurysm perforations. Of these six patients, the status of two deteriorated because of aneurysm perforation and another two because of thrombus formation (morbidity 3.6%). There were five deaths--one due to rerupture after embolization. Angiography follow-up studies of 87 aneurysms were obtained. Seventy aneurysms demonstrated progressive occlusion or a stable neck (80.5%), and 17 had some degree of recanalization (19.5%). The aneurysms originally diagnosed as a neck remnant showed a 15% rate of recanalization.
CONCLUSIONS: Matrix coils can be delivered into aneurysms with technical complications similar to those encountered using GDCs. Midterm anatomical outcomes to date have shown moderate improvement in the recanalization rate when compared with those realized using the GDC system. Because of the increased friction associated with the first-generation Matrix coil, the packing density in most aneurysms was less than that achieved with GDCs. Prolonged angiography follow-up evaluations are needed to document long-term efficacy.

Entities:  

Mesh:

Substances:

Year:  2006        PMID: 17219822     DOI: 10.3171/jns.2006.105.2.192

Source DB:  PubMed          Journal:  J Neurosurg        ISSN: 0022-3085            Impact factor:   5.115


  20 in total

1.  Bare, bio-active and hydrogel-coated coils for endovascular treatment of experimentally induced aneurysms. Long-term histological and scanning electron microscopy results.

Authors:  M H T Reinges; T Krings; A Y Drexler; A Ludolph; B Sellhaus; M Bovi; S Geibprasert; R Agid; K Scherer; F J Hans
Journal:  Interv Neuroradiol       Date:  2010-07-19       Impact factor: 1.610

2.  Matrix reloaded.

Authors:  H J Cloft
Journal:  AJNR Am J Neuroradiol       Date:  2007-10-09       Impact factor: 3.825

Review 3.  Materials and techniques for coiling of cerebral aneurysms: how much scientific evidence do we have?

Authors:  W Kurre; J Berkefeld
Journal:  Neuroradiology       Date:  2008-09-19       Impact factor: 2.804

4.  Giant intracranial aneurysms: evolution of management in a contemporary surgical series.

Authors:  Michael E Sughrue; David Saloner; Vitaliy L Rayz; Michael T Lawton
Journal:  Neurosurgery       Date:  2011-12       Impact factor: 4.654

5.  The natural history of "contrast stasis" within aneurysm after embolization.

Authors:  Mohamed M Abdihalim; Saqib A Chaudry; Nauman Tariq; M Fareed K Suri; Adnan I Qureshi
Journal:  J Vasc Interv Neurol       Date:  2013-02

6.  In memoriam: the matrix coil.

Authors:  W J van Rooij; M Sluzewski; J Peluso
Journal:  AJNR Am J Neuroradiol       Date:  2014-04-03       Impact factor: 3.825

7.  Gellan sulfate core platinum coil with tenascin-C promotes intra-aneurysmal organization in rats.

Authors:  Kazuhide Hamada; Yoichi Miura; Naoki Toma; Keiichi Miyamoto; Kyoko Imanaka-Yoshida; Satoshi Matsushima; Toshimichi Yoshida; Waro Taki; Hidenori Suzuki
Journal:  Transl Stroke Res       Date:  2014-06-19       Impact factor: 6.829

8.  Angiographic and histological comparison of canine bifurcation aneurysms treated with first generation matrix and standard GDC coils.

Authors:  Aquilla S Turk; Christopher M Luty; Victoria Carr-Brendel; Igor Polyakov; Dan Consigny; Julie Grinde; Rajat Mukherjee; Charles M Strother
Journal:  Neuroradiology       Date:  2007-09-27       Impact factor: 2.804

9.  Angiographic results in intracranial aneurysms treated with inert platinum coils.

Authors:  J R Vanzin; C Mounayer; D G Abud; R D'agostini Annes; J Moret
Journal:  Interv Neuroradiol       Date:  2012-12-03       Impact factor: 1.610

10.  Endovascular treatment of intracranial aneurysms with matrix detachable coils: midterm anatomic follow-up from a prospective multicenter registry.

Authors:  L Pierot; X Leclerc; A Bonafé; S Bracard
Journal:  AJNR Am J Neuroradiol       Date:  2007-10-05       Impact factor: 3.825

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.