| Literature DB >> 17215098 |
Fabiana S Felix1, Christopher M A Brett, Lúcio Angnes.
Abstract
Flow injection analysis (FIA) with amperometric detection was employed for acetaminophen quantification in pharmaceutical formulations using a carbon film resistor electrode. This sensor exhibited sharp and reproducible current peaks for acetaminophen without chemical modification of its surface. A wide linear working range (8.0x10(-7) to 5.0x10(-4) mol L(-1)) in phosphate buffer solution as well as high sensitivity (0.143 A mol(-1) L cm(-2)) and low submicromolar detection limit (1.36x10(-7) mol L(-1)) were achieved. The repeatability (R.S.D. for 10 successive injections of 5.0x10(-6) and 5.0x10(-5) mol L(-1) acetaminophen solutions) was 3.1 and 1.3%, respectively, without any memory effect between injections. The new procedure was applied to the analyses of commercial pharmaceutical products and the results were in good agreement with those obtained utilizing a spectrophotometric method. Consequently, this amperometric method has been shown to be very suitable for quality control analyses and other applications with similar requirements.Entities:
Mesh:
Substances:
Year: 2007 PMID: 17215098 DOI: 10.1016/j.jpba.2006.11.038
Source DB: PubMed Journal: J Pharm Biomed Anal ISSN: 0731-7085 Impact factor: 3.935