Estimation of tracheostomy tube cuff pressure by pilot balloon palpation.

C Faris1, E Koury, J Philpott, S Sharma, N Tolley, A Narula.   

Abstract

Two methods can be used to assess the intra-cuff pressure of tracheostomy tubes: digital palpation of the pilot balloon and use of a hand-held manometer. We conducted a telephone survey to determine the prevalence of both methods in intensive care units within 21 teaching hospitals across the United Kingdom. Forty-two per cent of the intensive care units surveyed used a protocol for monitoring cuff pressure with a manometer.A study to compare these two methods, using the manometer as the reference standard, was then carried out. The cuff pressure was correctly estimated in pre-inflated tracheostomy tubes, in a tracheal model, by 61 per cent of a cross-section of intensive care unit and otolaryngology staff.Using pilot balloon palpation is inaccurate and leaves a significant proportion of patients at risk of tracheal injury. We advocate the wider availability of hand-held pressure manometers in intensive care units and the institution of protocols for monitoring cuff pressure for any patient with a tracheostomy tube with an inflated cuff in situ.

Entities:  

Mesh:

Year:  2007        PMID: 17210092     DOI: 10.1017/S0022215107005324

Source DB:  PubMed          Journal:  J Laryngol Otol        ISSN: 0022-2151            Impact factor:   1.469


  2 in total

1.  Comparison of endotracheal tube cuff pressure values before and after training seminar.

Authors:  Ayça Tuba Dumanlı Özcan; Cihan Döğer; Abdülkadir But; Işık Kutlu; Şemsi Mustafa Aksoy
Journal:  J Clin Monit Comput       Date:  2017-07-22       Impact factor: 2.502

2.  Clinical or technological evaluation of depth of anesthesia at induction time: A simple study for an easy message.

Authors:  Péan Didier; Lejus Corinne
Journal:  Saudi J Anaesth       Date:  2013-01
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.