Literature DB >> 17207697

CT colonography: automatic measurement of polyp diameter compared with manual assessment - an in-vivo study.

D Burling1, S Halligan, S A Taylor, L Honeyfield, M E Roddie.   

Abstract

AIM: To investigate whether automated diameter assessment was feasible for CT colonography.
MATERIALS AND METHODS: Two experienced observers independently measured the maximum diameter of 50 polyps (colonoscopic reference size range 5-12 mm) from colonography datasets using conventionally placed software callipers and a variety of two-dimensional (2D) computed tomography (CT) window settings (colon, abdominal, bone, lung), and also three-dimensional (3D) perspective rendering. Polyps were also measured using automated polyp-segmentation software. Agreement between observers and with the colonoscopic reference measurement was determined using Bland-Altman, Wilcoxon, and Mann-Whitney U analyses.
RESULTS: Inter-observer agreement was similar for all window displays: mean difference in millimetres (SD difference; 95% limits of agreement) ranged from 0 (1.7, -3.3, 3.3) for 2D colon to -1.1mm (1.6, -4.3, 2.0) for 3D, compared with -0.5 (2.09, -4.6, 3.6) for automated measurement. When compared to colonoscopy, the largest discrepancy occurred using the 3D display (mean difference 1.3mm, 2.5mm for each observer). There was also a significant difference between estimates and reference size when using the 2D abdominal and 3D displays (p=0.03, <0.001).
CONCLUSION: Automated polyp measurement is possible in vivo. Automated and conventional methods have comparable inter-observer agreement. The greatest measurement error is encountered when using a 3D display for estimates of diameter.

Entities:  

Mesh:

Year:  2007        PMID: 17207697     DOI: 10.1016/j.crad.2006.09.018

Source DB:  PubMed          Journal:  Clin Radiol        ISSN: 0009-9260            Impact factor:   2.350


  4 in total

Review 1.  Polyp size measurement at CT colonography: what do we know and what do we need to know?

Authors:  Ronald M Summers
Journal:  Radiology       Date:  2010-06       Impact factor: 11.105

2.  Measurement of colonic polyps by radiologists and endoscopists: who is most accurate?

Authors:  S Punwani; S Halligan; P Irving; S Bloom; A Bungay; R Greenhalgh; J Godbold; S A Taylor; D G Altman
Journal:  Eur Radiol       Date:  2008-01-04       Impact factor: 5.315

3.  Linear measurement of polyps in CT colonography using level sets on 3D surfaces.

Authors:  Sovira Tan; Jianhua Yao; Michael M Ward; Ronald M Summers
Journal:  Conf Proc IEEE Eng Med Biol Soc       Date:  2009

4.  Variation of agreement in polyp size measurement between computed tomographic colonography and pathology assessment: clinical implications.

Authors:  Samir Gupta; Valerie Durkalski; Peter Cotton; Don C Rockey
Journal:  Clin Gastroenterol Hepatol       Date:  2008-02       Impact factor: 11.382

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.