Literature DB >> 17201839

Intrathoracic far-field electrocardiogram allows continuous monitoring of ischemia after total coronary occlusion.

Stefan Asbach1, Ingo Weiss, Beate Wenzel, Christoph Bode, Manfred Zehender.   

Abstract

INTRODUCTION: Recording of intrathoracic far-field electrocardiograms (FF-ECG) via can and electrodes of implantable cardioverter-defibrillators (ICD) is a promising method for continuous monitoring of myocardial ischemia. We assessed the hypothesis that experimentally induced ischemia provokes segment changes in the FF-ECG that can be detected by the ICD. METHODS AND
RESULTS: In seven pigs with an ICD implanted in the left pectoral region and electrodes placed in the right ventricle and the superior vena cava, we occluded all major coronary arteries in proximal and distal locations for 180 s each. Surface and FF-ECGs were compared for presence and time course of ischemic ST segment changes. Reliable detection of ischemia by ST segment analysis was possible in all (38/38) experiments. Maximum deviation from baseline was larger in FF-ECG (1.21 mV) than surface ECG leads (0.23 mV, P < 0.01) for all occlusion sites. Ischemia could be detected earlier (P < 0.05) in the FF-ECG, with a sensitivity of 100%, 93%, and 100% after occlusions in the left anterior descending, left circumflex, and right coronary arteries, respectively.
CONCLUSION: Intrathoracic FF-ECG allows reliable and reproducible detection of experimentally induced ischemia originating from all major coronary arteries and therefore could be an interesting tool for clinicians in monitoring high risk patients.

Entities:  

Mesh:

Year:  2006        PMID: 17201839     DOI: 10.1111/j.1540-8159.2006.00544.x

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  4 in total

1.  Subject of the year: remote monitoring.

Authors:  Fred Kusumoto; Nora Goldschlager
Journal:  J Interv Card Electrophysiol       Date:  2009-04-24       Impact factor: 1.900

Review 2.  Advances in remote monitoring of implantable pacemakers, cardioverter defibrillators and cardiac resynchronization therapy systems.

Authors:  Werner Jung; Andreas Rillig; Ralf Birkemeyer; Tomislav Miljak; Udo Meyerfeldt
Journal:  J Interv Card Electrophysiol       Date:  2008-09-27       Impact factor: 1.900

Review 3.  Remote monitoring of patients with implanted cardiac devices.

Authors:  Fred Kusumoto; Nora Goldschlager
Journal:  Clin Cardiol       Date:  2010-01       Impact factor: 2.882

4.  Continuous ST-Monitoring Function of Implantable Cardioverter Defibrillator Detects Silent Ischemia in Patients With Coronary Artery Disease.

Authors:  Tetsuya Watanabe; Keiji Hirooka; Yoshio Furukawa; Masanori Yabuki; Akio Hirata; Kazunori Kashiwase; Ryu Shutta; Takanao Mine; Hiroya Mizuno; Toshikazu Tanaka; Takahiro Doi; Akihiro Yoshida; Yuji Okuyama; Shinsuke Nanto
Journal:  J Am Heart Assoc       Date:  2018-06-30       Impact factor: 5.501

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.