Validity and repeatability of a new test for aniseikonia.

Beatriz Antona1, Francisco Barra, Ana Barrio, Enrique Gonzalez, Isabel Sanchez.   

Abstract

PURPOSE: The Aniseikonia Inspector 1.1 (AI) is a new software product to measure aniseikonia using red-green anaglyphs. The purpose of this study was to test whether the AI is a valid and reliable test.
METHODS: There were two groups of sample subjects: one at risk of aniseikonia, with anisometropia greater than or equal to 1.00 D (n= 29), and a control group (n= 45). The validity was studied by comparing the measured aniseikonia with the aniseikonia simulated with size lenses. The reliability was estimated by the Bland-Altman statistical method.
RESULTS: The results showed that the AI underestimated aniseikonia and that the underestimation was greater in the horizontal than in the vertical direction. The reliability was low, with biases that were clinically insignificant, but the 95% limits of agreement were around +/-2%. The behavior of the test was similar in both groups of subjects.
CONCLUSIONS: The reliability of the AI is only moderate, and professionals are therefore warned to use the results of this test with caution.

Mesh:

Year:  2007        PMID: 17197516     DOI: 10.1167/iovs.05-0575

Source DB:  PubMed          Journal:  Invest Ophthalmol Vis Sci        ISSN: 0146-0404            Impact factor:   4.799


  6 in total

1.  Measuring aniseikonia and investigating neuroplasticity and image factors in amblyopia (MAGNIFY): study protocol for a randomised clinical trial.

Authors:  Jayshree South; Tina Gao; Melinda Calderwood; Jason Turuwhenua; Paul Roberts; Arier Lee; Andrew Collins; Joanna Black
Journal:  Trials       Date:  2022-04-27       Impact factor: 2.728

2.  Aniseikonia Tests: The Role of Viewing Mode, Response Bias, and Size-Color Illusions.

Authors:  Miguel A García-Pérez; Eli Peli
Journal:  Transl Vis Sci Technol       Date:  2015-06-11       Impact factor: 3.283

3.  A systematic review of statistical methods used to test for reliability of medical instruments measuring continuous variables.

Authors:  Rafdzah Zaki; Awang Bulgiba; Noorhaire Nordin; Noor Azina Ismail
Journal:  Iran J Basic Med Sci       Date:  2013-06       Impact factor: 2.699

4.  The effects of optically and digitally simulated aniseikonia on stereopsis.

Authors:  David A Atchison; Thien Nguyen; Katrina L Schmid; Archayeeta Rakshit; Alex S Baldwin; Robert F Hess
Journal:  Ophthalmic Physiol Opt       Date:  2022-03-06       Impact factor: 3.992

5.  Clinical Aniseikonia in Anisometropia and Amblyopia.

Authors:  Jayshree South; Tina Gao; Andrew Collins; Arier Lee; Jason Turuwhenua; Joanna Black
Journal:  Br Ir Orthopt J       Date:  2020-11-20

6.  The relationship between clinical measures of aniseikonia and stereoacuity before and after LASIK.

Authors:  Ivana Mravicic; Maja Bohac; Selma Lukacevic; Kruno Jagaric; Merlak Maja; Sudi Patel
Journal:  J Optom       Date:  2019-10-25
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.