Literature DB >> 17197335

In vitro measurement of pressure differences using manometry at various injection speeds during discography.

Kwan-Sik Seo1, Richard Derby, Elaine S Date, Sang-Heon Lee, Byung-Jo Kim, Chang-Hyung Lee.   

Abstract

BACKGROUND CONTEXT: Lumbar discography has been widely used to evaluate discogenic low back pain. Anecdotal evidence suggests that pain reproduction during discography is more closely correlated with peak dynamic pressure than with static postinjection pressure. Although there can be a significant difference between dynamic and static pressures, to date most discographic evaluations use static pressure recorded postinjection (which is stable and easily measured). The use of readings taken after injection, rather than readings of maximum dynamic peak pressure recorded during injection, appear to increase false positives in lumbar discography. High-speed intradiscal injections also appear to have potentially confounding effects that may increase the rate of false-positive responses during lumbar discography. To date there has been no study for the evaluation of peak dynamic intradiscal pressures or for differentiating dynamic from static pressures in the nucleus pulposus (NP) in response to various speeds of intradiscal injection.
PURPOSE: The goal of this study was to obtain additional data on potential confounding factors that could affect discographic results by assessment of pressures within the NP during discography at various injection speeds. The purpose of data collection was to more precisely evaluate pressure differences between dynamic and static pressure within the NP, evident during discography. STUDY
DESIGN: In vitro laboratory study. SAMPLES: A total of 82 trials were performed on intervening discs of 82 porcine cadaver lumbar spines.
METHODS: Dynamic and static intradiscal pressures were measured with manometry, using two pressure sensors simultaneously during intradiscal injection of contrast media at various speeds. The tip of a 25-gauge needle was placed in the center of the NP and connected with a pressure manometer, which recorded the pressure and therefore the speed of injection. A second pressure reading was obtained using a sensor tip connected to a transducer; the sensor tip was located separately in the same NP tissue. The needle and transducer locations were confirmed by fluoroscopy. OUTCOME MEASURES: At low controlled injection speeds (below 0.08 mL/s), the mean peak pressure difference in the NP was 4.06 (+/-1.52) psi. With high-speed injections of 0.08 mL/s or greater, the mean increased abruptly up to 14.52 (+/-4.11) psi (p<.05). The data indicate that injections applied slowly resulted in smaller differences in pressure within the NP, registered by both manometry and the needle sensor tip. Additional samples were taken using both devices to confirm this threshold level.
RESULTS: With low injection speeds, especially those below 0.08 mL/s, differences between dynamic and static pressures on both pressure sensors were minor. These differences increased as injection speed became faster. However, at fast injection speeds of 0.08 mL/s and above, those differences were significantly higher.
CONCLUSION: Dynamic and static intradiscal pressures are of similar value when measured by manometer and by needle sensor at slow injection speeds during discography. However, the pressure differences appeared to rapidly increase in response to incremental increases in injection speed. The data from these 82 samples suggest that uncontrolled high speeds of intradiscal injections are a potential confounding factor, which may increase false-positive responses during lumbar discography.

Entities:  

Mesh:

Year:  2007        PMID: 17197335     DOI: 10.1016/j.spinee.2006.05.013

Source DB:  PubMed          Journal:  Spine J        ISSN: 1529-9430            Impact factor:   4.166


  6 in total

1.  HIZ's relation to axial load and low back pain: investigated with axial loaded MRI and pressure controlled discography.

Authors:  Hanna Hebelka; Hebelka Hanna; Tommy Hansson; Hansson Tommy
Journal:  Eur Spine J       Date:  2012-09-26       Impact factor: 3.134

Review 2.  Clinical diagnosis for discogenic low back pain.

Authors:  Yin-gang Zhang; Tuan-mao Guo; Xiong Guo; Shi-xun Wu
Journal:  Int J Biol Sci       Date:  2009-10-13       Impact factor: 6.580

3.  Automated pressure-controlled discography with constant injection speed and real-time pressure measurement.

Authors:  Hyoung Ihl Kim; Dong Ah Shin
Journal:  J Korean Neurosurg Soc       Date:  2009-07-31

4.  Clinical and radiological findings of discogenic low back pain confirmed by automated pressure-controlled discography.

Authors:  Hyung-Gon Kim; Dong-Ah Shin; Hyoung-Ihl Kim; Eun-Ae Yoo; Dong-Gyu Shin; Jung-Ok Lee
Journal:  J Korean Neurosurg Soc       Date:  2009-10-31

Review 5.  Provocative diskography: safety and predictive value in the outcome of spinal fusion or pain intervention for chronic low-back pain.

Authors:  Paul C Willems
Journal:  J Pain Res       Date:  2014-12-02       Impact factor: 3.133

Review 6.  The use of fiber Bragg grating sensors in biomechanics and rehabilitation applications: the state-of-the-art and ongoing research topics.

Authors:  Ebrahim Al-Fakih; Noor Azuan Abu Osman; Faisal Rafiq Mahamd Adikan
Journal:  Sensors (Basel)       Date:  2012-09-25       Impact factor: 3.576

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.