Literature DB >> 17197148

Automated determination of flutamide by a validated flow-injection method: application to dissolution studies of pharmaceutical tablets.

Paraskevas D Tzanavaras1, Demetrius G Themelis.   

Abstract

The first flow-injection (FI) method for the determination of flutamide--a potent antiandrogen used for the treatment of prostate cancer--is reported. The method is based on the direct measurement of the absorbance of the analyte at 310 nm under flow conditions. Parameters affecting the determination such as detection wavelength, sample injection volume and flow rate were studied and optimized. The assay was validated (linearity, limits of detection and quantitation, accuracy, repeatability, reproducibility and selectivity) for the dissolution studies of flutamide-containing tablets during stability testing. The results were in good agreement with high performance liquid chromatography (HPLC) used as a reference method.

Entities:  

Mesh:

Substances:

Year:  2006        PMID: 17197148     DOI: 10.1016/j.jpba.2006.11.039

Source DB:  PubMed          Journal:  J Pharm Biomed Anal        ISSN: 0731-7085            Impact factor:   3.935


  2 in total

1.  Resonance Rayleigh scattering, second-order scattering and frequency doubling scattering spectra of copper(II)-Flutamide system with anionic surfactants and its analytical application.

Authors:  Ayaz Mohd; Aftab Aslam Parwaz Khan; Shaista Bano; K S Siddiqi
Journal:  J Fluoresc       Date:  2011-01-11       Impact factor: 2.217

2.  Determination of flutamide and two major metabolites using HPLC-DAD and HPTLC methods.

Authors:  Nada S Abdelwahab; Heba A H Elshemy; Nehal F Farid
Journal:  Chem Cent J       Date:  2018-01-25       Impact factor: 4.215