Initial experience with the Cardiva Boomerang vascular closure device in diagnostic catheterization.

Brendan J Doyle1, Michael J Godfrey, Ryan J Lennon, James L Ryan, John F Bresnahan, Charanjit S Rihal, Henry H Ting.   

Abstract

OBJECTIVES: The authors studied the safety and efficacy of the Cardiva Boomerang vascular closure device in patients undergoing diagnostic cardiac catheterization.
BACKGROUND: Conventional vascular closure devices (sutures, collagen plugs, or metal clips) have been associated with catastrophic complications including arterial occlusion and foreign body infections; furthermore, they cannot be utilized in patients with peripheral vascular disease or vascular access site in a vessel other than the common femoral artery. The Cardiva Boomerang device facilitates vascular hemostasis without leaving any foreign body behind at the access site, can be used in peripheral vascular disease, and can be used in vessels other than the common femoral artery
METHODS: A total of 96 patients undergoing transfemoral diagnostic cardiac catheterization were included in this study, including 25 (26%) patients with contraindications to conventional closure devices. Femoral angiography was performed prior to deployment of the Cardiva Boomerang closure device. Patients were ambulated at 1 hr after hemostasis was achieved.
RESULTS: The device was successfully deployed and hemostasis achieved with the device alone in 95 (99%) patients. The device failed to deploy in 1 (1%) patient and required conversion to standard manual compression. Minor complications were observed in 5 (5%) patients. No patients experienced major complications including femoral hematoma > 4 cm, red blood cell transfusion, retroperitoneal bleed, arteriovenous fistula, pseudoaneurysm, infection, arterial occlusion, or vascular surgery.
CONCLUSIONS: The Cardiva Boomerang device is safe and effective in patients undergoing diagnostic cardiac catheterization using the transfemoral approach, facilitating early ambulation with low rates of vascular complications. (c) 2006 Wiley-Liss, Inc.

Entities:  

Mesh:

Year:  2007        PMID: 17195965     DOI: 10.1002/ccd.20937

Source DB:  PubMed          Journal:  Catheter Cardiovasc Interv        ISSN: 1522-1946            Impact factor:   2.692


  2 in total

1.  Successful use of the Cardiva Boomerang™ vascular closure device to close a brachial artery puncture site after emergency PTCA.

Authors:  Plinio Cirillo; Gianluca Petrillo; Greta Luana D'Ascoli; Federico Piscione; Massimo Chiariello
Journal:  Heart Vessels       Date:  2010-09-28       Impact factor: 2.037

2.  Usefulness of a Nonsuture Closure Device in Patients Undergoing Diagnostic Coronary and Peripheral Angiography.

Authors:  John T Owens; Shaun Bhatty; Robert J Donovan; Andrea Tordini; Peter Danyi; Kalpesh Patel; Jacob A Wegelin; Ion S Jovin
Journal:  Int J Angiol       Date:  2017-10-04