Accuracy of the apex locating function of TCM Endo V in simulated conditions: a comparison study.

Ozgür Topuz1, Ozgür Uzun, Ali Cemal Tinaz, Burak Sadik.   

Abstract

OBJECTIVE: The aim of this study was to evaluate the accuracy of the apex-locating function of the TCM Endo V and to compare the results to these of the Root ZX. STUDY
DESIGN: Forty-seven extracted maxillary incisor teeth with mature apices and patent single root canals were used for the study. After the root surfaces were cleaned and access cavities were prepared, the true working lengths (TL) were determined with a #10 stainless steel file. All samples were mounted in an alginate model especially developed to test the apex locators. Both electronic apex locators were used according to the manufacturer's recommendations and adjusted to 0.5-mm accuracy to the apical foramen. EL(1) was designates as the working length measured with Root ZX and EL(2) was designated as the working length measured with TCM Endo V. A paired t test was used to statistically analyze the significance of the mean differences at the 95% confidence interval of the differences (P = .001) among TL, EL(1), and EL(2).
RESULTS: There were no statistically significant differences between TL and EL(1), TL and EL(2), and EL(1) and EL(2) (P = .001). Both Root ZX and TCM Endo V gave similar results.
CONCLUSIONS: TCM Endo V proved to be as reliable as Root ZX but the use of the device to determine the working length was not easy as Root ZX. Further studies are needed to compare TCM Endo V in clinical conditions with similar electric motor and apex locator combined devices.

Entities:  

Mesh:

Year:  2006        PMID: 17188532     DOI: 10.1016/j.tripleo.2006.09.004

Source DB:  PubMed          Journal:  Oral Surg Oral Med Oral Pathol Oral Radiol Endod        ISSN: 1079-2104


  4 in total

1.  Influence of Open Apex on Working Length Determination Using Cone-Beam Computed Tomography and Apex Locators: A Comparative In Vitro Study.

Authors:  Swati Srivastava; Hanan M Alharbi; Mai Soliman; Elzahraa Eldwakhly; Manal M Abdelhafeez
Journal:  Biomed Res Int       Date:  2022-07-23       Impact factor: 3.246

2.  An in vitro stereomicroscopic comparative evaluation of a combination of apex locator and endodontic motor with an integrated endodontic motor.

Authors:  Ch Swarupa; Girija S Sajjan; Yv Sashi Kanth
Journal:  J Conserv Dent       Date:  2013-09

3.  Simulated root resorption: a new study model.

Authors:  Jamileh Ghoddusi; Saeed Asgary; Masoud Parirokh; Mohammad Jafar Eghbal; Mahdi Vatanpour; Fatemeh Shahrami
Journal:  Iran Endod J       Date:  2008-01-10

4.  Comparative evaluation of the accuracy of two electronic apex locators in determining the working length in teeth with simulated apical root resorption: An in vitro study.

Authors:  Vidya Saraswathi; Archit Kedia; Tina Puthen Purayil; Vasudev Ballal; Aakriti Saini
Journal:  J Conserv Dent       Date:  2016 Sep-Oct
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.