Abstract
This study evaluated a new immunochromatographic assay (SAS Legionella Test) for its ability to detect Legionella pneumophila serogroup 1 antigen in urine. Results were compared with those obtained using the Binax Now urinary antigen test. Sensitivity and specificity were estimated as 82.9% and 99.0%, respectively, for the SAS Legionella Test, and 91.4% and 100%, respectively, for the Binax Now urinary antigen test. The sensitivity of both tests increased to 97.1% (p 0.009) and 94.2% (p 0.7), respectively, if the tests were examined after 1 h.Entities:
Mesh:
Substances:
Year: 2007 PMID: 17184292 DOI: 10.1111/j.1469-0691.2006.01587.x
Source DB: PubMed Journal: Clin Microbiol Infect ISSN: 1198-743X Impact factor: 8.067