3D thoracoscopic ultrasound volume measurement validation in an ex vivo and in vivo porcine model of lung tumours.

V D M Hornblower1, E Yu, A Fenster, J J Battista, R A Malthaner.   

Abstract

The purpose of this study was to validate the accuracy and reliability of volume measurements obtained using three-dimensional (3D) thoracoscopic ultrasound (US) imaging. Artificial "tumours" were created by injecting a liquid agar mixture into spherical moulds of known volume. Once solidified, the "tumours" were implanted into the lung tissue in both a porcine lung sample ex vivo and a surgical porcine model in vivo. 3D US images were created by mechanically rotating the thoracoscopic ultrasound probe about its long axis while the transducer was maintained in close contact with the tissue. Volume measurements were made by one observer using the ultrasound images and a manual-radial segmentation technique and these were compared with the known volumes of the agar. In vitro measurements had average accuracy and precision of 4.76% and 1.77%, respectively; in vivo measurements had average accuracy and precision of 8.18% and 1.75%, respectively. The 3D thoracoscopic ultrasound can be used to accurately and reproducibly measure "tumour" volumes both in vivo and ex vivo.

Entities:  

Mesh:

Substances:

Year:  2006        PMID: 17183130     DOI: 10.1088/0031-9155/52/1/007

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  2 in total

1.  Measurement of liquid and solid component parameters in canine vocal fold lamina propria.

Authors:  Robert Phillips; Yu Zhang; Megan Keuler; Chao Tao; Jack J Jiang
Journal:  J Acoust Soc Am       Date:  2009-04       Impact factor: 1.840

2.  Human Cadaveric Artificial Lung Tumor-Mimic Training Model.

Authors:  Réka Székely; Ferenc Imre Suhai; Kinga Karlinger; Gábor Baksa; Bence Szabaczki; László Bárány; Gergely Pölöskei; Gergely Rácz; Ödön Wagner; Béla Merkely; Tamás Ruttkay
Journal:  Pathol Oncol Res       Date:  2021-04-26       Impact factor: 3.201

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.