Literature DB >> 17181255

A microfluidic channel flow cell for electrochemical ESR.

Andrew J Wain1, Richard G Compton, Rudolph Le Roux, Sinéad Matthews, Kamran Yunus, Adrian C Fisher.   

Abstract

The design, fabrication, and characterization of microfluidic channel flow devices for in situ simultaneous hydrodynamic electrochemical ESR is reported. The microelectrochemical reactors consist of gold film electrodes situated within rectangular ducts of height 350 microm and widths in the range 500-2000 microm. The small dimensions of the channels result in minimal dielectric loss when centralized within a cylindrical TE011 resonant cavity, leading to a high level of sensitivity. This is demonstrated by using the one-electron oxidation of N,N,N',N'-tetramethyl-p-phenylenediamine (TMPD) in acetonitrile as a model system, wherein the ESR spectra obtained for the corresponding stable radical cation are of a high signal-to-noise ratio. Signal intensity is measured as a function of flow rate for this system, and the behavior is validated by means of 3-dimensional numerical modeling of the hydrodynamic flow profile.

Entities:  

Year:  2006        PMID: 17181255     DOI: 10.1021/jp064903b

Source DB:  PubMed          Journal:  J Phys Chem B        ISSN: 1520-5207            Impact factor:   2.991


  1 in total

1.  XAS and EPR in Situ Observation of Ru(V) Oxo Intermediate in a Ru Water Oxidation Complex.

Authors:  Natalia Levin; Carla Casadevall; George E Cutsail; Julio Lloret-Fillol; Serena DeBeer; Olaf Rüdiger
Journal:  ChemElectroChem       Date:  2021-12-27       Impact factor: 4.782

  1 in total