Comparison of the color of natural teeth measured by a colorimeter and Shade Vision System.

Byeong-Hoon Cho1, Yong-Kyu Lim, Yong-Keun Lee.   

Abstract

OBJECTIVES: The objectives were to measure the difference in the color and color parameters of natural teeth measured by a tristimulus colorimeter (CM, used as a reference) and Shade Vision System (SV), and to determine the influence of color parameters on the color difference between the values measured by two instruments.
METHOD: Color of 12 maxillary and mandibular anterior teeth was measured by CM and SV for 47 volunteers (number of teeth=564). Color parameters such as CIE L*, a* and b* values, chroma and hue angle measured by two instruments were compared. Chroma was calculated as C*ab=(a*2 = b*2)1/2, and hue angle was calculated as h degrees =arctan(b*/a*). The influence of color parameters measured by CM on the color difference (DeltaE*(ab)) between the values measured by two instruments was analyzed with multiple regression analysis (alpha=0.01).
RESULTS: Mean DeltaE*(ab) value between the values measured by two instruments was 21.7 (+/-3.7), and the mean difference in lightness (CIE L*) and chroma was 16.2 (+/-3.9) and 13.2 (+/-3.0), respectively. Difference in hue angle was high as 132.7 (+/-53.3) degrees . Except for the hue angle, all the color parameters showed significant correlations and the coefficient of determination (r(2)) was in the range of 0.089-0.478. Based on multiple regression analysis, the standardized partial correlation coefficient (beta) of the included predictors for the color difference was -0.710 for CIE L* and -0.300 for C*(ab) (p<0.01).
CONCLUSION: All the color parameters showed significant but weak correlations except for hue angle. When lightness and chroma of teeth were high, color difference between the values measured by two instruments was small. Clinical accuracy of two instruments should be investigated further.

Entities:  

Mesh:

Year:  2006        PMID: 17178151     DOI: 10.1016/j.dental.2006.11.008

Source DB:  PubMed          Journal:  Dent Mater        ISSN: 0109-5641            Impact factor:   5.304


  4 in total

1.  Could digital imaging be an alternative for digital colorimeters?

Authors:  Alper Caglar; Kivanc Yamanel; Kamran Gulsahi; Bora Bagis; Mutlu Ozcan
Journal:  Clin Oral Investig       Date:  2009-08-18       Impact factor: 3.573

2.  An in Vitro Comparison of Coronal Discolouration Caused by White Mineral Trioxide Aggregate, Theracal, Calcium-Enriched Mixture and Biodentine.

Authors:  Zohreh Khalilak; Ehsan Esnaashari; Keivan Saati; Delaram Bineshmarvasti; Hazhir Yousefshahi; Mahsima Nobakht
Journal:  Eur Endod J       Date:  2022-03

3.  Lightness, chroma, and hue distributions in natural teeth measured by a spectrophotometer.

Authors:  Teuta Pustina-Krasniqi; Kujtim Shala; Gloria Staka; Teuta Bicaj; Enis Ahmedi; Linda Dula
Journal:  Eur J Dent       Date:  2017 Jan-Mar

4.  Color comparison between non-vital and vital teeth.

Authors:  Delia Cristina Greta; Horaţiu Alexandru Colosi; Cristina Gasparik; Diana Dudea
Journal:  J Adv Prosthodont       Date:  2018-06-12       Impact factor: 1.904

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.