Literature DB >> 17174960

Unsupervised parameter optimization for automated retention time alignment of severely shifted gas chromatographic data using the piecewise alignment algorithm.

Karisa M Pierce1, Bob W Wright, Robert E Synovec.   

Abstract

Simulated chromatographic separations were used to study the performance of piecewise retention time alignment and to demonstrate automated unsupervised (without a training set) parameter optimization. The average correlation coefficient between the target chromatogram and all remaining chromatograms in the data set was used to optimize the alignment parameters. This approach frees the user from providing class information and makes the alignment algorithm applicable to classifying completely unknown data sets. The average peak in the raw simulated data set was shifted up to two peak-widths-at-base (average relative shift=2.0) and after alignment the average relative shift was improved to 0.3. Piecewise alignment was applied to severely shifted GC separations of gasolines and reformate distillation fraction samples. The average relative shifts in the raw gasolines and reformates data were 4.7 and 1.5, respectively, but after alignment improved to 0.5 and 0.4, respectively. The effect of piecewise alignment on peak heights and peak areas is also reported. The average relative difference in peak height was -0.20%. The average absolute relative difference in area was 0.15%.

Mesh:

Substances:

Year:  2006        PMID: 17174960     DOI: 10.1016/j.chroma.2006.11.101

Source DB:  PubMed          Journal:  J Chromatogr A        ISSN: 0021-9673            Impact factor:   4.759


  3 in total

1.  Comprehensive two-dimensional gas chromatography/time-of-flight mass spectrometry peak sorting algorithm.

Authors:  Cheolhwan Oh; Xiaodong Huang; Fred E Regnier; Charles Buck; Xiang Zhang
Journal:  J Chromatogr A       Date:  2007-12-03       Impact factor: 4.759

2.  Automatic time-shift alignment method for chromatographic data analysis.

Authors:  Qing-Xia Zheng; Hai-Yan Fu; He-Dong Li; Bing Wang; Cui-Hua Peng; Sheng Wang; Jun-Lan Cai; Shao-Feng Liu; Xiao-Bing Zhang; Yong-Jie Yu
Journal:  Sci Rep       Date:  2017-03-21       Impact factor: 4.379

3.  Combining peak- and chromatogram-based retention time alignment algorithms for multiple chromatography-mass spectrometry datasets.

Authors:  Nils Hoffmann; Matthias Keck; Heiko Neuweger; Mathias Wilhelm; Petra Högy; Karsten Niehaus; Jens Stoye
Journal:  BMC Bioinformatics       Date:  2012-08-27       Impact factor: 3.169

  3 in total