Correction factor in Orbscan II in the assessment of corneal pachymetry.

Arthur C K Cheng1, Emily Tang, Shaheeda Mohamed, Dennis S C Lam.   

Abstract

PURPOSE: Comparison of corneal pachymetry assessment with ultrasound and Orbscan II using acoustic factor and subtraction methods.
METHODS: Ultrasound and Orbscan pachymetry for all patients with LASIK performed between July 2002 and May 2003 were retrospectively analyzed. Comparison between the 2 preoperative measurements was analyzed.
RESULTS: Using a custom acoustic factor of 0.93, there was no significant difference between the overall ultrasound and Orbscan pachymetry (P = 0.696). However, there was underestimation in thick corneas and overestimation in thin corneas. Using the subtraction method as the correcting technique, the magnitude of over- and underestimation was reduced.
CONCLUSION: The current correction method by means of acoustic factor may result in error in assessing extremes of corneal thickness. Alternative approaches such as the subtraction method can be considered to achieve more accurate results.

Entities:  

Mesh:

Year:  2006        PMID: 17172890     DOI: 10.1097/01.ico.0000240089.26467.dc

Source DB:  PubMed          Journal:  Cornea        ISSN: 0277-3740            Impact factor:   2.651


  2 in total

1.  Central Corneal Thickness Measurement Using Ultrasonic Pachymetry, Rotating Scheimpflug Camera, and Scanning-slit Topography Exclusively in Thin Non-keratoconic Corneas.

Authors:  Mehrdad Mohammadpour; Kazem Mohammad; Nasser Karimi
Journal:  J Ophthalmic Vis Res       Date:  2016 Jul-Sep

2.  Effect of post crosslinking haze on the repeatability of Scheimpflug-based and slit-scanning imaging devices.

Authors:  Rohit Shetty; Aarti Agrawal; Rashmi Deshmukh; Luci Kaweri; Harsha L Rao; Harsha Nagaraja; Chaitra Jayadev
Journal:  Indian J Ophthalmol       Date:  2017-04       Impact factor: 1.848

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.