| Literature DB >> 17169823 |
Michelle Gill1, Gavin R Graff, Alan J Adler, Raed A Dweik.
Abstract
We tested reproducibility of exhaled nitric oxide (FE(NO)) and inter-operator handling when measured with a handheld device, NIOX MINO. We enrolled 20 volunteers using a priori goals of acceptable reproducibility to be mean within-subject standard deviation less than 3 parts per billion (ppb) for FE(NO) measurements less than 30 ppb, and mean coefficient of variation less than 10% for FE(NO) measurements more than 30 ppb. Seventeen subjects with measurements less than 30 ppb displayed a mean standard deviation of 1.15, and 3 subjects with FE(NO) more than 30 ppb had a mean coefficient of variation of 2.4%. We conclude that NIOX MINO demonstrates excellent reproducibility for all ranges of FE(NO).Entities:
Mesh:
Substances:
Year: 2006 PMID: 17169823 DOI: 10.1080/02770900601031045
Source DB: PubMed Journal: J Asthma ISSN: 0277-0903 Impact factor: 2.515