Real-time detection and alerting for acute ST-segment elevation myocardial ischemia using an implantable, high-fidelity, intracardiac electrogram monitoring system with long-range telemetry in an ambulatory porcine model.

Tim A Fischell1, David R Fischell, Robert E Fischell, Renu Virmani, Jennifer J DeVries, Mitchell W Krucoff.   

Abstract

OBJECTIVES: The purpose of this study was to evaluate feasibility of using real-time, high-fidelity, intracardiac electrogram monitoring from a permanently implantable ischemia detection system (IIDS), with long-range telemetry capability to detect ST-segment shifts associated with acute or subacute coronary occlusion in a porcine model.
BACKGROUND: Early identification of coronary occlusion with ST-segment elevation could profoundly accelerate the timing of revascularization and improve clinical outcomes.
METHODS: This paper reports the first investigation using real-time, high-fidelity, intracardiac electrogram monitoring from a permanently IIDS, with long-range telemetry capability. This IIDS was tested in an ambulatory porcine model, with acute coronary occlusion precipitated by stent thrombosis. Two overlapping copper stents were implanted in the left anterior descending (n = 3), the circumflex (n = 3), or the right coronary artery (n = 2) of juvenile farm pigs. Monitoring was carried using telemetry from the IIDS.
RESULTS: All stented pigs had acute ST-segment elevation event(s) triggering the alerting thresholds of the IIDS. All triggered events were confirmed to be caused by thrombosis of the copper stent(s), and well correlated to infarct age and location. Four of the 8 pigs died from ventricular fibrillation, recorded by the IIDS at a mean time of 70 +/- 121 h after ST-segment alert. The sensitivity and specificity of alerting for ST-segment shift, associated with thrombotic coronary occlusion, were 100% and 100%, respectively.
CONCLUSIONS: This study demonstrates the ability of an implantable ischemia detection system to detect ST-segment elevation from coronary occlusion in a porcine model of ST-segment elevation myocardial infarction. ST-segment elevation was sufficient to trigger alerting thresholds in all 3 epicardial coronary distributions. Such a system, with real-time alerting capability, could advance the time frame of reperfusion therapy and potentially prevent, rather than interrupt, acute myocardial infarction in patients with coronary artery disease.

Entities:  

Mesh:

Year:  2006        PMID: 17161264     DOI: 10.1016/j.jacc.2006.07.058

Source DB:  PubMed          Journal:  J Am Coll Cardiol        ISSN: 0735-1097            Impact factor:   24.094


  4 in total

1.  A statistically based acute ischemia detection algorithm suitable for an implantable device.

Authors:  Bruce Hopenfeld; M Sasha John; Tim A Fischell; Steven R Johnson
Journal:  Ann Biomed Eng       Date:  2012-06-28       Impact factor: 3.934

Review 2.  Remote monitoring of patients with implanted cardiac devices.

Authors:  Fred Kusumoto; Nora Goldschlager
Journal:  Clin Cardiol       Date:  2010-01       Impact factor: 2.882

3.  Clinical valuation of ST changes in a group of patients with ventricular arrhythmias: The inSighT Study.

Authors:  Jerzy Krzysztof Wranicz; Michał Kałowski; Dirk Bastian; Aparna Jaswal; Christof Kolb; Edgar Zitron; Iwona Cygankiewicz; Krzysztof Kaczmarek
Journal:  Ann Noninvasive Electrocardiol       Date:  2022-02-15       Impact factor: 1.485

4.  Continuous ST-Monitoring Function of Implantable Cardioverter Defibrillator Detects Silent Ischemia in Patients With Coronary Artery Disease.

Authors:  Tetsuya Watanabe; Keiji Hirooka; Yoshio Furukawa; Masanori Yabuki; Akio Hirata; Kazunori Kashiwase; Ryu Shutta; Takanao Mine; Hiroya Mizuno; Toshikazu Tanaka; Takahiro Doi; Akihiro Yoshida; Yuji Okuyama; Shinsuke Nanto
Journal:  J Am Heart Assoc       Date:  2018-06-30       Impact factor: 5.501

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.