| Literature DB >> 17161028 |
James D Daly1, Colleen M O'Hehir, George M Frame.
Abstract
A method for measurement of an important biological marker, 1,1'-sulfonylbis[2-(methylthio)ethane] (SBMTE) of sulfur mustard agent HD [bis-(2-chloroethyl)sulfide] in human urine, to quantify HD exposure, is presented. It employs TiCl3 reduction of beta-lyase metabolites to SBMTE, and automated solid-phase extraction sample preparation, followed by isotope dilution liquid chromatography-positive ion-electrospray ionization-tandem mass spectrometry with 7.5 min/sample cycle time, to achieve SBMTE quantitation of up to 200 samples/24h a day. Percent relative standard deviations over the calibration range varied from 12.0% at 0.1 ng/mL to 0.9% at 100 ng/mL, and the limit of detection from a 0.5 mL sample was below the lowest level calibration standard of 0.1 ng/mL.Entities:
Mesh:
Substances:
Year: 2006 PMID: 17161028 DOI: 10.1016/j.jchromb.2006.11.031
Source DB: PubMed Journal: J Chromatogr B Analyt Technol Biomed Life Sci ISSN: 1570-0232 Impact factor: 3.205