A method for the routine quantitative determination of the major serotonin metabolite 5-hydroxyindole-3-acetic acid (5-HIAA) in urine is described. 5-HIAA was analyzed without prior sample cleanup, using an automated high-performance liquid chromatography system with isocratic elution and electrochemical detection (+0.60 V versus a Ag/AgCl reference electrode). The urine samples were mixed with a solution of the internal standard (5-hydroxyindole-3-propionic acid) and centrifuged. The supernatant was transferred to sealed glass vials, and a 2-microliters aliquot was injected directly onto a C18 reversed-phase analytical column, using an automatic sample injector. Samples of urine could be stored for several months at -80 or at +7 degrees C for 2 days without loss of 5-HIAA. However, a gradual decline with time occurred in crude samples stored at room temperature or above, as well as in urine samples diluted with the mobile phase. The detector response was linear in the range of 0-65 mumol/l 5-HIAA, and the intra- and interassay coefficients of variation were about 5 and 7%, respectively (n = 10).Entities: