Joshua G Bales1, Jimmie Colon, Usha Ramadhyani, Elma LeDoux, James T Bennett.
Abstract
Pediatric heart condition management may include a variety of implanted cardiac devices. Monopolar electrocautery (Bovie) produces significant electromagnetic interference to these devices. This interference can alter the function of a cardiac generator resulting in a variety of complications including impaired cardiac output and asystole. We report the case of a 16 plus 2-year-old girl who presented with idiopathic scoliosis and a past medical history significant for complete congenital heart block treated with a DDD*** pacemaker. During surgery, the use monopolar electrocautery caused the patient to develop asystole and loss of pulsatile blood pressure. Electrocautery was subsequently terminated, the patient then returned to a normal sinus rhythm and blood pressure normalized.Entities: