Literature DB >> 17154704

Drug-eluting and bare nitinol stents for the treatment of atherosclerotic lesions in the superficial femoral artery: long-term results from the SIROCCO trial.

Stephan H Duda1, Marc Bosiers, Johannes Lammer, Dierk Scheinert, Thomas Zeller, Vincent Oliva, Alexander Tielbeek, John Anderson, Benjamin Wiesinger, Gunnar Tepe, Alexandra Lansky, Michael R Jaff, Catharina Mudde, Hans Tielemans, Jean-Paul Beregi.   

Abstract

PURPOSE: To review clinical outcomes of patients with chronic limb ischemia and TASC type C lesions treated with sirolimus-eluting versus bare SMART nitinol self-expanding stents.
METHODS: Data were obtained from a randomized, multicenter, double-blinded study conducted in 2 phases. All 93 patients had chronic limb ischemia and superficial femoral artery (SFA) occlusions or stenoses (average lesion length 8.3 cm). In total, 47 patients (31 men; mean age 66.3+/-9.1 years, range 50-84) received the sirolimus-eluting SMART stent and 46 patients (36 men; mean age 65.9 +/-10.8 years, range 38-83) received a bare SMART nitinol stent. Both groups were followed for a mean 24 months.
RESULTS: Both the sirolimus-eluting and the bare SMART stents were effective in revascularizing the diseased SFA and in sustaining freedom from restenosis. For both types of stents, improvements in ankle-brachial indices (ABI) and symptoms of claudication were maintained over 24 months (median 24-month ABI 0.96 for the sirolimus group versus 0.87 for the bare stent group, p>0.05). At 24 months, the restenosis rate in the sirolimus group was 22.9% versus 21.1% in the bare stent group (p>0.05). The cumulative in-stent restenosis rates according to duplex ultrasound were 4.7%, 9.0%, 15.6%, and 21.9%, respectively, at 6, 9, 18, and 24 months; the rates did not differ significantly between the treatment groups. The TLR rate for the sirolimus group was 6% and for the bare stent group 13%; the TVR rates were somewhat higher: 13% and 22%, respectively. Mortality rates did not differ significantly between the groups.
CONCLUSION: These data demonstrate that the sirolimus-eluting and the bare SMART stent are effective, safe, and free from restenosis in a majority of patients for up to 24 months. Because the restenosis rate in the bare stent group is unexpectedly low, no significant difference could be found between the sirolimus-eluting and the bare SMART stents.

Entities:  

Mesh:

Substances:

Year:  2006        PMID: 17154704     DOI: 10.1583/05-1704.1

Source DB:  PubMed          Journal:  J Endovasc Ther        ISSN: 1526-6028            Impact factor:   3.487


  67 in total

1.  The consequences of the mechanical environment of peripheral arteries for nitinol stenting.

Authors:  Michael Early; Daniel J Kelly
Journal:  Med Biol Eng Comput       Date:  2011-08-11       Impact factor: 2.602

Review 2.  Current state of diagnosis and management of critical limb ischemia.

Authors:  Dan Clair; Samir Shah; John Weber
Journal:  Curr Cardiol Rep       Date:  2012-04       Impact factor: 2.931

3.  Arterial geometry, flow pattern, wall shear and mass transport: potential physiological significance.

Authors:  G Coppola; C Caro
Journal:  J R Soc Interface       Date:  2008-11-25       Impact factor: 4.118

Review 4.  Magnetically targeted delivery of therapeutic agents to injured blood vessels for prevention of in-stent restenosis.

Authors:  Michael Chorny; Ilia Fishbein; Richard F Adamo; Scott P Forbes; Zoë Folchman-Wagner; Ivan S Alferiev
Journal:  Methodist Debakey Cardiovasc J       Date:  2012-01

Review 5.  [Peripheral vascular disease of iliac and femoro-popliteal arteries: state-of-the-art endoluminal revascularization].

Authors:  A Chavan; L Luthe; B Schmuck
Journal:  Radiologe       Date:  2010-01       Impact factor: 0.635

Review 6.  Cost-effectiveness analysis of drug-coated therapies in the superficial femoral artery.

Authors:  Natalie D Sridharan; Aureline Boitet; Kenneth Smith; Kathy Noorbakhsh; Efthymios Avgerinos; Mohammad H Eslami; Michel Makaroun; Rabih Chaer
Journal:  J Vasc Surg       Date:  2017-09-27       Impact factor: 4.268

Review 7.  Drug-coated balloon and stent therapies for endovascular treatment of atherosclerotic superficial femoral artery disease.

Authors:  Subhash Banerjee; Karan Sarode; Atif Mohammad; Emmanouil S Brilakis
Journal:  Curr Cardiol Rep       Date:  2015-05       Impact factor: 2.931

Review 8.  Treating femoropopliteal disease: established and emerging technologies.

Authors:  Athanasios Diamantopoulos; Konstantinos Katsanos
Journal:  Semin Intervent Radiol       Date:  2014-12       Impact factor: 1.513

Review 9.  [New stent developments for peripheral arterial occlusive disease].

Authors:  G Tepe
Journal:  Radiologe       Date:  2010-01       Impact factor: 0.635

10.  Current approach to the diagnosis and treatment of femoral-popliteal arterial disease. A systematic review.

Authors:  Christos Kasapis; Hitinder S Gurm
Journal:  Curr Cardiol Rev       Date:  2009-11
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.