Balloon-assisted coiling through a 5-French system.

J Bradley White1, Kennith F Layton, David F Kallmes, Harry J Cloft.   

Abstract

INTRODUCTION: We present a catheter technique that utilizes a 5F system for the purpose of balloon-assisted coiling in the setting of intracranial aneurysms.
METHODS: A standard 5F short sheath is placed in the common femoral artery, and a 5F diagnostic catheter is placed through the sheath and used for selective vessel angiography. When endovascular intervention is pursued, the diagnostic catheter is placed in the appropriate vessel and systemic heparinization is ensured. Over an exchange length wire, the 5F vertebral catheter and 5F short sheath are exchanged for a 5F Shuttle (Cook) sheath. We then routinely place a 10, 14 or 18 microcatheter over an appropriately gauged microguidewire into the aneurysm. As needed, balloon catheters are then placed across the neck of the aneurysm for remodeling purposes. During the course of the procedure, control angiography is performed through the Shuttle sheath. Following the placement of coils, the microcatheter and balloon catheter are removed and a final biplane image is obtained via the 5F Shuttle sheath.
RESULTS: This technique has been employed in 15 patients who required balloon-assisted coiling of an intracranial aneurysm. There were no technical difficulties or arterial access site complications from the procedures. Catheter mobility and torque were not affected, nor was the quality of our imaging.
CONCLUSION: We conclude that this small-diameter system provides ample "room" for catheter placement and interventional treatment while reducing the known risks of postprocedural complications. Angiographic images remain excellent and are comparable to those obtained by larger catheters.

Entities:  

Mesh:

Year:  2006        PMID: 17146636     DOI: 10.1007/s00234-006-0163-3

Source DB:  PubMed          Journal:  Neuroradiology        ISSN: 0028-3940            Impact factor:   2.804


  4 in total

1.  Balloon-assisted coiling through a single 6F guiding catheter.

Authors:  G D Luzardo; I B Ross; G Gal
Journal:  AJNR Am J Neuroradiol       Date:  2006-01       Impact factor: 3.825

2.  Study of the performance of 5 French and 7 French catheters in coronary angiography: a functional comparison.

Authors:  R S Kohli; G W Vetrovec; S A Lewis; S Cole
Journal:  Cathet Cardiovasc Diagn       Date:  1989-11

3.  Anatomic and clinical factors associated with complications of transfemoral arteriography.

Authors:  M P Lilly; W Reichman; A A Sarazen; W I Carney
Journal:  Ann Vasc Surg       Date:  1990-05       Impact factor: 1.466

Review 4.  Current treatment methods for postcatheterization pseudoaneurysms.

Authors:  Robert Morgan; Anna-Maria Belli
Journal:  J Vasc Interv Radiol       Date:  2003-06       Impact factor: 3.464

  4 in total
  5 in total

1.  Is the risk of balloon assistance underinflated?

Authors:  M Sluzewski; W J van Rooij; Harry Cloft
Journal:  AJNR Am J Neuroradiol       Date:  2008-09-03       Impact factor: 3.825

2.  Balloon-assisted rapid intermittent sequential coiling (BRISC) technique for the treatment of complex wide-necked intracranial aneurysms.

Authors:  J Modi; M Eesa; B K Menon; J H Wong; M Goyal
Journal:  Interv Neuroradiol       Date:  2011-04-29       Impact factor: 1.610

3.  Balloon-assisted coiling of intracranial aneurysms is not associated with a higher complication rate.

Authors:  Boris Lubicz; Florence Lefranc; Michaël Bruneau; Danielle Balériaux; Olivier De Witte
Journal:  Neuroradiology       Date:  2008-05-14       Impact factor: 2.804

4.  A 6-Fr guiding catheter (Slim Guide(®)) for use with multiple microdevices.

Authors:  Yutaka Kai; Yuki Ohmori; Masaki Watanabe; Yasuyuki Kaku; Motohiro Morioka; Teruyuki Hirano; Shigetoshi Yano; Takayuki Kawano; Jun-Ichiro Hamada; Jun-Ichi Kuratsu
Journal:  Surg Neurol Int       Date:  2012-06-09

5.  Double-balloon Trapping for Coil Embolization of Ruptured Internal Carotid Artery Aneurysm: A Novel Technique.

Authors:  Ahmed Ansari; Tomotaka Ohshima; Shunsaku Goto; Taiki Yamamoto; Kojiro Ishikawa; Yoko Kato
Journal:  Asian J Neurosurg       Date:  2019 Jul-Sep
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.