The mechanical properties of recovered PMMA bone cement: a preliminary study.

R P S Chaplin1, A J C Lee, R M Hooper, M Clarke.   

Abstract

Samples of polymethylmethacrylate (PMMA) bone cement, used in the fixation of hip prostheses, have been recovered from 11 patients after in service life spans of between 15 and 24 years. Eighteen samples in total have been recovered from the acetabular and/or femoral cement. Samples were subjected to three point bending tests, their density, porosity and microhardness determined and all specimens were examined using EDX and X-ray techniques. Since the porosity of many of the samples is very high, the continuous matrix properties are inferred from the performance of individual specimens. No evidence has been found to suggest that the PMMA has deteriorated whilst in-vivo and the mechanical properties of the cement matrices appear to be comparable to freshly made PMMA.

Entities:  

Mesh:

Substances:

Year:  2006        PMID: 17143776     DOI: 10.1007/s10856-006-0619-3

Source DB:  PubMed          Journal:  J Mater Sci Mater Med        ISSN: 0957-4530            Impact factor:   3.896


  7 in total

Review 1.  Key issues involved with the use of miniature specimens in the characterization of the mechanical behavior of polymeric biomaterials--a review.

Authors:  Gladius Lewis
Journal:  J Biomed Mater Res       Date:  2002

2.  [New PMMA bone cements for vacuum mixing systems].

Authors:  K Schelling; C Heisel; S M Schnürer; H Mau; S J Breusch
Journal:  Orthopade       Date:  2002-06       Impact factor: 1.087

3.  The initiation of failure in cemented femoral components of hip arthroplasties.

Authors:  M Jasty; W J Maloney; C R Bragdon; D O O'Connor; T Haire; W H Harris
Journal:  J Bone Joint Surg Br       Date:  1991-07

4.  Migration of metal and polyethylene particles from articular prostheses may generate lymphadenopathy with histiocytosis.

Authors:  M F Baslé; G Bertrand; S Guyetant; D Chappard; M Lesourd
Journal:  J Biomed Mater Res       Date:  1996-02

5.  Early aseptic loosening of uncemented femoral components in primary total hip replacement. A review based on the Norwegian Arthroplasty Register.

Authors:  L I Havelin; B Espehaug; S E Vollset; L B Engesaeter
Journal:  J Bone Joint Surg Br       Date:  1995-01

Review 6.  [Bone cements based on polymethylmethacrylate].

Authors:  S J Breusch; K-D Kühn
Journal:  Orthopade       Date:  2003-01       Impact factor: 1.087

Review 7.  The science of bone cement: a historical review.

Authors:  Frank R DiMaio
Journal:  Orthopedics       Date:  2002-12       Impact factor: 1.390

  7 in total
  2 in total

1.  Ultrastructural evaluation of in vitro mineralized calcium phosphate phase on surface phosphorylated poly(hydroxy ethyl methacrylate-co-methyl methacrylate).

Authors:  G S Sailaja; P Ramesh; H K Varma
Journal:  J Mater Sci Mater Med       Date:  2010-01-23       Impact factor: 3.896

2.  The long-term in vivo behavior of polymethyl methacrylate bone cement in total hip arthroplasty.

Authors:  Hiroyuki Oonishi; Haruhiko Akiyama; Mitsuru Takemoto; Toshiyuki Kawai; Koji Yamamoto; Takao Yamamuro; Hironobu Oonishi; Takashi Nakamura
Journal:  Acta Orthop       Date:  2011-10       Impact factor: 3.717

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.