New monitoring software for larger clinical application of brachial artery flow-mediated vasodilatation measurements.

Damian Craiem1, Gilles Chironi, Jerome Gariepy, Jennifer Miranda-Lacet, Jaime Levenson, Alain Simon.   

Abstract

BACKGROUND: The reproducibility of brachial artery flow-mediated vasodilatation (FMD) is limited by the operator dependence of most measurement methods.
METHODS: A new automated computerized analysis of brachial artery ultrasound scan providing a continuous evolution of the diameter during acute hyperemia, reactive to short hyperemia of the forearm and hand, was tested in 10 normal volunteers and 26 asymptomatic patients with cardiovascular risk factors such as hypertension, hypercholesterolemia, heavy smoking, history of premature coronary heart disease and the metabolic syndrome. FMD was the percentage of the maximum hyperemic diastolic diameter from baseline. Within-reading variations in FMD and diameters were assessed by reading one scan from the same subject twice by two observers. The within-subject variability of FMD was assessed by analysing two repeated measurements in the same subject by the same operator 1 h, 1 week or 1 month apart.
RESULTS: Coefficients of variation (CV) of repeated FMD readings were 7.5% in normal volunteers and 6.9% in patients with risk factors. CV of repeated FMD measurements 1 h apart were 7.8% in normal volunteers and 16.5% in patients with risk factors. In normal volunteers, CV of repeated FMD measurements 1 week apart was 9.6%, and in patients with risk factors CV of repeated FMD measurement 1 month apart was 18.1%.
CONCLUSION: This method overcomes the variability of FMD measurement seen with conventional manual analysis in normal volunteers, and to a lesser extent in patients with major cardiovascular risk factors, thus supporting its clinical applicability to patients with disease conditions.

Entities:  

Mesh:

Year:  2007        PMID: 17143184     DOI: 10.1097/HJH.0b013e3280109287

Source DB:  PubMed          Journal:  J Hypertens        ISSN: 0263-6352            Impact factor:   4.844


  4 in total

1.  Linear and nonlinear viscoelastic modeling of aorta and carotid pressure-area dynamics under in vivo and ex vivo conditions.

Authors:  Daniela Valdez-Jasso; Daniel Bia; Yanina Zócalo; Ricardo L Armentano; Mansoor A Haider; Mette S Olufsen
Journal:  Ann Biomed Eng       Date:  2011-01-04       Impact factor: 3.934

2.  Reliability of the passive leg movement assessment of vascular function in men.

Authors:  H Jonathan Groot; Ryan M Broxterman; Jayson R Gifford; Ryan S Garten; Matthew J Rossman; Catherine L Jarrett; Oh Sung Kwon; Jay R Hydren; Russell S Richardson
Journal:  Exp Physiol       Date:  2022-04-05       Impact factor: 2.858

3.  The area under curve for time-course analysis parameters is associated with abdominal aortic aneurysms and the severity of peripheral artery disease in men.

Authors:  Nao Konno; Taku Harada; Daijirou Akamatsu; Hitoshi Goto; Takashi Miki; Takashi Kamei; Masahiro Kohzuki
Journal:  Int J Cardiol Hypertens       Date:  2021-02-04

4.  Model-based analysis of flow-mediated dilation and intima-media thickness.

Authors:  G Bartoli; G Menegaz; M Lisi; G Di Stolfo; S Dragoni; T Gori
Journal:  Int J Biomed Imaging       Date:  2009-04-06
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.