Atsushi Hirohata1, Yoshihiro Morino, Junya Ako, Ryota Sakurai, Maurice Buchbinder, Ronald P Caputo, Steven P Karas, Gregory J Mishkel, Michael R Mooney, Charles D O'shaughnessy, Albert E Raizner, Robert L Wilensky, David O Williams, Shing-Chiu Wong, Paul G Yock, Yasuhiro Honda, Jeffrey W Moses, Peter J Fitzgerald.
Abstract
A direct coronary stenting technique using drug-eluting stents may decrease drug-eluting stent efficacy due to possible damage to the surface coating of the stent. The DIRECT is a multicenter, prospective, nonrandomized trial designed to evaluate the direct stenting strategy for the sirolimus-eluting Bx-Velocity stent compared with the historical control (SIRIUS trial, stenting with predilation). Volumetric and cross-sectional intravascular ultrasound analyses at 8-month follow-up were performed in 115 patients (DIRECT n= 64, control n = 51). Patient and lesion characteristics were comparable between groups. The DIRECT group achieved an equivalent uniform expansion index, defined as minimum stent area/maximum stent area x 100, compared with the control group (65.9 +/- 11.7 vs 63.1 +/- 12.7, p = NS). At 8-month follow-up, vessel, stent, lumen, and neointimal volume index (volume in cubic millimeters/length in millimeters) and percent neointimal volume were similar between the DIRECT and control groups (vessel volume index 13.9 +/- 4.40 vs 15.0 +/- 3.83; stent volume index 6.83 +/- 2.02 vs 6.94 +/- 2.04; lumen volume index 6.71 +/- 2.04 vs 6.81 +/- 2.07; neointimal volume index 0.14 +/- 0.24 vs 0.16 +/- 0.23; percent neointimal volume 3.73 +/- 6.97 vs 3.14 +/- 5.32, p = NS for all). In addition, in-stent neointimal hyperplasia distribution was significantly smaller near the distal stent edge (0.22 vs 0.098 mm(3)/mm, p = 0.01 for an average neointimal volume index within 3 mm from the distal stent edge). In conclusion, direct coronary stenting with the sirolimus-eluting Bx-Velocity stent is equally effective in terms of uniform stent expansion and long-term quantitative intravascular ultrasound results compared with conventional stenting using predilation. This strategy appears to be associated with less neointimal hyperplasia near the distal stent edge.Entities:
Mesh:
Substances:
Year: 2006 PMID: 17126651 DOI: 10.1016/j.amjcard.2006.06.046
Source DB: PubMed Journal: Am J Cardiol ISSN: 0002-9149 Impact factor: 2.778