Literature DB >> 17119584

Phase-step calibration technique based on a two-run-times-two-frame phase-shift method.

Xianghong Zhong1.   

Abstract

A novel phase-step calibration technique is presented on the basis of a two-run-times-two-frame phase-shift method. First the symmetry factor M is defined to describe the distribution property of the distorted phase due to phase-shifter miscalibration; then the phase-step calibration technique, in which two sets of two interferograms with a straight fringe pattern are recorded and the phase step is obtained by calculating M of the wrapped phase map, is developed. With this technique, a good mirror is required, but no uniform illumination is needed and no complex mathematical operation is involved. This technique can be carried out in situ and is applicable to any phase shifter, whether linear or nonlinear.

Year:  2006        PMID: 17119584     DOI: 10.1364/ao.45.008863

Source DB:  PubMed          Journal:  Appl Opt        ISSN: 1559-128X            Impact factor:   1.980


  1 in total

1.  Piezoelectric Actuated Phase Shifter Based on External Laser Interferometer: Design, Control and Experimental Validation.

Authors:  Peng-Zhi Li; Xiao-Dong Wang; Yong-Xin Sui; De-Fu Zhang; Dong-Fang Wang; Li-Jian Dong; Ming-Yang Ni
Journal:  Sensors (Basel)       Date:  2017-04-11       Impact factor: 3.576

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.