Literature DB >> 17118315

Evaluation of the surface radiation dose and dose gradient in early stage breast cancer using high-dose-rate brachytherapy MammoSite applicator.

Amir Sadeghi1, Bradley Prestidge, Jui-Min Lee, Arthur Rosenthal.   

Abstract

PURPOSE: To monitor the radiation dose delivered and dose homogeneity to the skin and adjacent tissues around a MammoSite applicator. METHODS AND MATERIALS: Radiation surface dose was measured on 67 consecutive patients with Stage T1-T2N0M0 breast cancer implanted with the MammoSite balloon. Thermoluminescent dosimeters and metal oxide semiconductor field effect transistors were used to measure surface dose directly over the balloon, contralateral breast, thyroid, axilla, and inframammary fold. The dose homogeneity was retrospectively studied from the treatment plans in 25 patients.
RESULTS: The mean maximum skin dose was 267 cGy, with a range of 56-488 cGy per fraction, 4 cGy (+/-2 cGy) to the contralateral breast, 8 cGy (+/-4 cGy) to the thyroid, 47 cGy to the ipsilateral axilla (+/-33 cGy), and 52 cGy (+/-47 cGy) to the ipsilateral inframmammary fold. The mean dose gradients were 127%, 110%, 99%, 96%, and 89% for inflated volumes of 35, 40, 50, 60, and 70 mL, respectively.
CONCLUSIONS: All patients completed therapy with no or minimal skin reaction with radiation doses comparable to or better to external beam radiation. Considering the dose homogeneity, the larger inflated volume allows for better homogeneity of dose distribution, minimizes the volume of contiguous tissue that receives relatively high doses, and may reduce the risk of fat necrosis.

Entities:  

Mesh:

Year:  2006        PMID: 17118315     DOI: 10.1016/j.brachy.2006.02.004

Source DB:  PubMed          Journal:  Brachytherapy        ISSN: 1538-4721            Impact factor:   2.362


  4 in total

1.  Determination of exit skin dose for 192Ir intracavitary accelerated partial breast irradiation with thermoluminescent dosimeters.

Authors:  Julie A Raffi; Stephen D Davis; Cliff G Hammer; John A Micka; Keith A Kunugi; Jana E Musgrove; John W Winston; Terresa J Ricci-Ott; Larry A DeWerd
Journal:  Med Phys       Date:  2010-06       Impact factor: 4.071

Review 2.  Present state and issues in IORT Physics.

Authors:  Frank W Hensley
Journal:  Radiat Oncol       Date:  2017-01-27       Impact factor: 3.481

3.  Brachytherapy in accelerated partial breast irradiation (APBI) - review of treatment methods.

Authors:  Janusz Skowronek; Magdalena Wawrzyniak-Hojczyk; Kinga Ambrochowicz
Journal:  J Contemp Brachytherapy       Date:  2012-09-29

4.  Dose estimation for different skin models in interstitial breast brachytherapy.

Authors:  Judyta Lasota; Renata Kabacińska; Roman Makarewicz
Journal:  J Contemp Brachytherapy       Date:  2014-06-03
  4 in total