Effect of 7.0 Tesla MRI on upper eyelid implants.

Thomas Schrom1, Ariane Thelen, Patrick Asbach, Hans-Christian Bauknecht.   

Abstract

To ensure the MRI compatibility of various eyelid implants in high-field MRI, 3 eyelid weights made of pure gold (99.99%), pure platinum (99.95%), and a platinum (97%)/iridium (3%) alloy were examined in vitro. Temperature changes, position changes, and imaging artifacts of the different implants were determined in a small-bore 7.0 Tesla MRI system. The 7.0 Tesla MRI system demonstrated that none of the eyelid implants carried a risk of heating or dislocation; therefore, these implants are MRI compatible up to a magnetic field strength of 7.0 Tesla.

Entities:  

Mesh:

Substances:

Year:  2006        PMID: 17117110     DOI: 10.1097/01.iop.0000240807.42320.d9

Source DB:  PubMed          Journal:  Ophthalmic Plast Reconstr Surg        ISSN: 0740-9303            Impact factor:   1.746


  1 in total

1.  Evaluation of 39 medical implants at 7.0 T.

Authors:  David X Feng; Joseph P McCauley; Fea K Morgan-Curtis; Redoan A Salam; David R Pennell; Mary E Loveless; Adrienne N Dula
Journal:  Br J Radiol       Date:  2015-10-20       Impact factor: 3.039