Thomas Schrom1, Ariane Thelen, Patrick Asbach, Hans-Christian Bauknecht.
Abstract
To ensure the MRI compatibility of various eyelid implants in high-field MRI, 3 eyelid weights made of pure gold (99.99%), pure platinum (99.95%), and a platinum (97%)/iridium (3%) alloy were examined in vitro. Temperature changes, position changes, and imaging artifacts of the different implants were determined in a small-bore 7.0 Tesla MRI system. The 7.0 Tesla MRI system demonstrated that none of the eyelid implants carried a risk of heating or dislocation; therefore, these implants are MRI compatible up to a magnetic field strength of 7.0 Tesla.Entities:
Mesh:
Substances:
Year: 2006 PMID: 17117110 DOI: 10.1097/01.iop.0000240807.42320.d9
Source DB: PubMed Journal: Ophthalmic Plast Reconstr Surg ISSN: 0740-9303 Impact factor: 1.746