New corticosteroid-eluting porous polyethylene implant for the management of lower eyelid retraction: a pilot study.

Daniel C Garibaldi1, Michael R Robinson, Susan S Lee, Dong Jun J Park, Howard F Fine, Luke Deitz, Sam D'Anna, Peng Yuan, Nicholas Ranson, Michael P Grant, Nicholas T Iliff, Shannath L Merbs.   

Abstract

PURPOSE: Lower eyelid retraction after trauma presents a challenging management problem. We postulated that a porous polyethylene (pPE) eyelid spacer coated with a polyvinyl alcohol (PVA) and triamcinolone acetonide (TA) matrix could deliver corticosteroid locally over extended periods and modulate inflammation and scar formation. We designed a pPE corticosteroid-eluting implant and evaluated its characteristics in vitro and in vivo.
METHODS: The release characteristics of pPE implants coated with a PVA/TA matrix of low, intermediate, and high doses of TA were studied in vitro. The implants were then placed in the posterior lamella of lower eyelids of Dutch Belted rabbits for 12 weeks. Clinical events were recorded and eyelids were examined for gross and histologic features, including capsular thickness and degree of vascularity, fibrovascular ingrowth, and inflammatory response.
RESULTS: In vitro, implants coated with the intermediate and high doses of TA released the drug at a steady rate for at least 78 days. In rabbits, the PVA and PVA/TA coating prevented fibrovascular ingrowth, except where breaks in the PVA/TA coat were present. Implants with PVA/TA coating demonstrated less inflammation and capsule vascularity. An inverse correlation between TA dose and capsule thickness was noted.
CONCLUSIONS: We describe a novel drug-release pPE eyelid implant. The corticosteroid-eluting implant demonstrated antiangiogenic and anti-inflammatory properties, which could prove beneficial in the treatment of lower eyelid retraction.

Entities:  

Mesh:

Substances:

Year:  2006        PMID: 17117094     DOI: 10.1097/01.iop.0000245487.61479.56

Source DB:  PubMed          Journal:  Ophthalmic Plast Reconstr Surg        ISSN: 0740-9303            Impact factor:   1.746


  2 in total

1.  Blood Collection within the Maxillary Sinus following Orbital Fracture Repair: The Impact of Mesh Implants and Drains.

Authors:  Jacob T Cox; Jing Tian; Shannath Merbs; Nicholas R Mahoney
Journal:  Craniomaxillofac Trauma Reconstr       Date:  2019-01-16

Review 2.  Current concepts of ocular adnexal surgery.

Authors:  Maria Borrelli; Gerd Geerling
Journal:  GMS Interdiscip Plast Reconstr Surg DGPW       Date:  2013-02-27
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.